Keysight U2741A Programmer’s Reference Guide
[SENSe:]RESistance:NPLC
Syntax
[SENSe:]RESistance:NPLC <PLC|MIN|MAX|DEF>
This command sets the integration time in number of power line cycles (PLCs) for
resistance measurements. Integration time affects the measurement resolution
(for better resolution, use a longer integration time) and measurement speed (for
faster measurements, use a shorter integration time).
[SENSe:]RESistance:NPLC?
This command queries the NPLC settings for resistance measurements.
Parameter
Remarks
– Only those integration times set to an integral number of power line cycles
(1, 2, 10, or 20 PLCs) provides a normal mode (line frequency noise) rejection.
– The U2741A sets the integration time to 20 PLCs after a factory reset (
*RST
command) or an Instrument Preset (
SYST:PRES
command).
Examples
RES:NPLC 1
//Sets the NPLC for resistance
measurements to 1 PLC
RES:NPLC?
//Queries the NPLC setting for
resistance measurements
Typical Response:
+1.00
See Also
Item
Type
Range of values
Default value
PLC
Numeric
{0|0.025|1|2|10|20}
MIN = 0 PLC
MAX = 20 PLC
20