STATus Subsystem
7
Keysight U2741A Programmer’s Reference Guide
117
Remarks
– The following table lists the bit definitions for the Standard Operation Register.
– Use the
<enable_value>
parameter to specify which bits will be reported to
the Status Byte. The decimal value specified corresponds to the
binary-weighted sum of the bits you wish to enable in the register. For
example, to enable bit 5 (decimal value = 32) and bit 8 (decimal value = 256),
the corresponding decimal value would be 288 (32 + 256).
– The clear status (
*CLS
) command will not clear the enable register but it does
clear all the bits in the event register.
– The
STAT:PRES
command will clear all the bits in the enable register.
– The
*RST
command has no effect on this register.
Examples
STAT:OPER:ENAB 32
//Sets the enable register (bit 5 is set)
STAT:OPER:ENAB?
//Queries the enable register
Typical Response:
+32
See Also
Bit number
Decimal value
Definition
0
Calibration in Progress
1
U2741A is performing a calibration
1 to 3
Not Used
Not Used
0 is returned
4
Measuring
16
U2741A is initiated, and is making, or about to
make a measurement
5
Waiting for Trigger
32
U2741A is waiting for a trigger
6 to 7
Not Used
Not Used
0 is returned
8
Configuration Change
256
U2741A configuration has been changed
9
Not Used
Not Used
0 is returned
10
Not Used
Not Used
0 is returned
11 to 15
Not Used
Not Used
0 is returned
Summary of Contents for U2741A
Page 1: ...Keysight U2741A USB Modular 5 5 Digits Digital Multimeter Programmer s Reference Guide ...
Page 8: ...8 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 10: ...10 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 12: ...12 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...