![Keysight U2741A Programmer'S Reference Manual Download Page 103](http://html.mh-extra.com/html/keysight/u2741a/u2741a_programmers-reference-manual_1961514103.webp)
SENSe Subsystem
6
Keysight U2741A Programmer’s Reference Guide
103
[SENSe:]VOLTage:AC:RANGe
Syntax
[SENSe:]VOLTage:AC:RANGe <range|MIN|MAX|AUTO>
This command sets the measurement range for AC voltage measurements.
[SENSe:]VOLTage:AC:RANGe?
This command queries the measurement range for AC voltage measurements.
Parameter
Remarks
– Selecting a discrete range will disable autorange.
– The U2741A is set to 10 V, with autorange enabled, after a factory reset (
*RST
command)
Examples
VOLT:AC:RANG 1
//Configures the U2741A to measure AC
voltage at 1 V range
VOLT:AC:RANG?
//Queries the voltage range
Typical Response:
+1.00
See Also
Item
Type
Range of values
Default value
range
Numeric
100 mV, 1 V, 10 V, 100 V, 1 kV
MIN = 100 mV
MAX = 1 kV
10 V