![Keysight U2741A Programmer'S Reference Manual Download Page 45](http://html.mh-extra.com/html/keysight/u2741a/u2741a_programmers-reference-manual_1961514045.webp)
CONFigure Subsystem
4
Keysight U2741A Programmer’s Reference Guide
45
CONFigure:FREQuency
Syntax
CONFigure:FREQuency[<range|MIN|MAX|DEF>
[,<resolution|MIN|MAX|DEF>]]
This command presets and configures the U2741A for frequency measurements
with the specified range and resolution. This command does not initiate the
measurement.
Parameters
Remarks
– The
CONF
command does not place the U2741A in the “wait-for-trigger” state.
Use the
INIT
or
READ?
command in conjunction with
CONF
to place the
U2741A in the “wait-for-trigger” state.
– Frequency is measured on AC voltage signals. Thus, the proper AC voltage
should be first set using the
[SENS:]VOLT:AC:RANG
command. The voltage
range should be 0.1 V or greater for more accurate frequency measurement.
– The range parameter is required only to specify the measurement resolution
parameter. Therefore, it is not necessary to send a new command for each new
frequency to be measured.
– If the input signal voltage is greater than can be measured on the selected
voltage range (manual range), the U2741A gives an overload indication of
+9.9E+37
from the remote interface. Autorange can be enabled for the input
voltage.
– The frequency measurement does not autorange.
– With no signal applied,
0
is returned.
Item
Type
Range of values
Default value
range
Numeric
1 Hz, 1 MHz
1 Hz
resolution
Numeric
N/A
5.5
Summary of Contents for U2741A
Page 1: ...Keysight U2741A USB Modular 5 5 Digits Digital Multimeter Programmer s Reference Guide ...
Page 8: ...8 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 10: ...10 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 12: ...12 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...