7
STATus Subsystem
118
Keysight U2741A Programmer’s Reference Guide
STATus:OPERation[:EVENt]?
Syntax
STATus:OPERation[:EVENt]?
This command queries the event register for the Standard Operation Register
group. This is a read-only register and the bits are cleared when you read the
register.
Remarks
– The condition register bits reflect the current condition. If a condition goes
away, the corresponding bit is cleared in the condition register.
– The following table lists the bit definitions for the Standard Operation Register.
– Once a bit is set, it remains set until cleared by reading the event register or
the clear status (
*CLS
) command.
– The
*RST
command has no effect on this register.
NOTE
For more information on the SCPI Status System, see
.
Bit number
Decimal value
Definition
0
Calibration in Progress
1
U2741A is performing a calibration
1 to 3
Not Used
Not Used
0 is returned
4
Measuring
16
U2741A is initiated, and is making, or about to make a
measurement
5
Waiting for Trigger
32
U2741A is waiting for a trigger
6 to 7
Not Used
Not Used
0 is returned
8
Configuration Change
256
U2741A configuration has been changed
9
Not Used
Not Used
0 is returned
10
Not Used
Not Used
0 is returned
11 to 15
Not Used
Not Used
0 is returned
Summary of Contents for U2741A
Page 1: ...Keysight U2741A USB Modular 5 5 Digits Digital Multimeter Programmer s Reference Guide ...
Page 8: ...8 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 10: ...10 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 12: ...12 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...