![Keysight U2741A Programmer'S Reference Manual Download Page 139](http://html.mh-extra.com/html/keysight/u2741a/u2741a_programmers-reference-manual_1961514139.webp)
Other Commands
10
Keysight U2741A Programmer’s Reference Guide
139
FETCh?
Syntax
FETCh?
This command transfers the reading to the U2741A output buffer where you can
read it into your PC.
Remarks
– The
FETCh?
command will wait until the measurement is complete to
terminate.
– The U2741A clears any reading from volatile memory after a factory reset
(
*RST
command) when mainframe power is cycled, when the configuration is
changed (see
Chapter 4, “CONFigure Subsystem”
), or on an
INIT
command.
Examples
CONF 10, 0.00003
//Configures to measure DC voltage at
10 V range and 30
μ
V resolution
TRIG:SOUR IMM
//Sets the trigger source to immediate
INIT
//Places the U2741A in
wait-for-trigger state
FETC?
//Transfers the reading from memory to
the U2741A output buffer
Typical Response:
+1.00