10
Other Commands
140
Keysight U2741A Programmer’s Reference Guide
INITiate[:IMMediate]
Syntax
INITiate[:IMMediate]
This command changes the state of the triggering system from the “idle” state to
the “wait-for-trigger” state. Measurements will begin when the specified trigger
conditions are satisfied following the receipt of the
INIT
command.
Remarks
– To retrieve the reading, use the
FETC?
command.
– The
ABOR
command may be used to set the U2741A to idle state.
– The
INIT
command only initiates the measurement and needs a trigger (
BUS
or
IMMediate
) to make the actual measurement.
Examples
CONF 10, 0.00003
//Configures to measure DC voltage at
10 V range and 30
μ
V resolution
TRIG:SOUR IMM
//Sets the trigger source to immediate
INIT
//Places the U2741A in
wait-for-trigger state
FETC?
//The FETCh? command transfers the
reading from memory to the U2741A
output buffer
Typical Response:
+1.00
See Also
NOTE
The INIT command also clears the previous reading from the memory.