12
SCPI Error Messages
162
Keysight U2741A Programmer’s Reference Guide
Device Dependent Errors
The table below contains the device-dependent errors.
Table 12-3
Device Dependent Errors
(these errors set Standard Event Status register bit #3)
Error number
Device dependent errors
–310
System error
This indicates that some error, termed as “system error” by the device, has occurred. The U2741A internal
operation has not completed properly, possibly due to an abnormal hardware or firmware condition.
–313
Calibration memory lost
The nonvolatile memory used to store the U2741A calibration constants has detected a checksum error which
may be caused by a device failure or extreme conditions such as lightning or strong magnetic fields
–315
Configuration memory lost
The nonvolatile memory used to store the U2741A configuration settings has detected a checksum error which
may be caused by a device failure or extreme conditions such as lightning or strong magnetic fields
–321
Out of memory
There is insufficient working memory available
–330
Self-test failed
An executed self-test failed
–350
Queue overflow
The error queue is full because more than 20 errors have occurred. No additional errors are stored until you
remove the errors from the queue.
Summary of Contents for U2741A
Page 1: ...Keysight U2741A USB Modular 5 5 Digits Digital Multimeter Programmer s Reference Guide ...
Page 8: ...8 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 10: ...10 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 12: ...12 Keysight U2741A Programmer s Reference Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...