:MEASure Commands
21
Keysight P9241/42/43A Oscilloscopes SCPI Programmer's Guide
457
:MEASure:OVERshoot
Command Syntax
:MEASure:OVERshoot [<source>]
<source> ::= {CHANnel<n> | FUNCtion<m> | MATH<m> | WMEMory<r>}
<n> ::= 1 to (# analog channels) in NR1 format
<m> ::= 1 to (# math functions) in NR1 format
<r> ::= 1 to (# ref waveforms) in NR1 format
The :MEASure:OVERshoot command installs a screen measurement and starts an
overshoot measurement. If the optional source parameter is specified, the current
source is modified.
Query Syntax
:MEASure:OVERshoot? [<source>]
The :MEASure:OVERshoot? query measures and returns the overshoot of the edge
closest to the trigger reference, displayed on the screen. The method used to
determine overshoot is to make three different vertical value measurements: Vtop,
Vbase, and either Vmax or Vmin, depending on whether the edge is rising or
falling.
For a rising edge:
overshoot = ((Vmax-Vtop) / (Vtop-Vbase)) x 100
For a falling edge:
overshoot = ((Vbase-Vmin) / (Vtop-Vbase)) x 100
Vtop and Vbase are taken from the normal histogram of all waveform vertical
values. The extremum of Vmax or Vmin is taken from the waveform interval right
after the chosen edge, halfway to the next edge. This more restricted definition is
used instead of the normal one, because it is conceivable that a signal may have
more preshoot than overshoot, and the normal extremum would then be
dominated by the preshoot of the following edge.
Return Format
<overshoot><NL>
<overshoot>::= the percent of the overshoot of the selected waveform in
NR3 format
See Also
•
"Introduction to :MEASure Commands"
•
•
NOTE
This command is not available if the source is FFT (Fast Fourier Transform).
Summary of Contents for P9241
Page 1: ...Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Page 6: ...6 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Page 36: ...36 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 1 What s New...
Page 41: ...Setting Up 2 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 41...
Page 44: ...44 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 2 Setting Up...
Page 222: ...222 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 7 Root Commands...
Page 244: ...244 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands...
Page 256: ...256 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands...
Page 278: ...278 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 10 CHANnel n Commands...
Page 328: ...328 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands...
Page 386: ...386 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands...
Page 390: ...390 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands...
Page 416: ...416 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands...
Page 498: ...498 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 21 MEASure Commands...
Page 532: ...532 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 22 MTESt Commands...
Page 572: ...572 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 24 SAVE Commands...
Page 900: ...900 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 28 TIMebase Commands...
Page 1080: ...1080 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 31 WGEN w Commands...
Page 1142: ...1142 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 34 Error Messages...
Page 1344: ...1344 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide Index...