:FUNCtion<m> Commands
17
Keysight P9241/42/43A Oscilloscopes SCPI Programmer's Guide
355
:FUNCtion<m>[:FFT]:CENTer
Command Syntax
:FUNCtion<m>[:FFT]:CENTer <frequency>
<m> ::= 1 to (# math functions) in NR1 format
<frequency> ::= the current center frequency in NR3 format.
The range
of legal values is from -25 GHz to 25 GHz.
The :FUNCtion<m>[:FFT]:CENTer command sets the center frequency when FFT
(Fast Fourier Transform) is selected.
Query Syntax
:FUNCtion<m>[:FFT]:CENTer?
The :FUNCtion<m>[:FFT]:CENTer? query returns the current center frequency in
Hertz.
Return Format
<frequency><NL>
<frequency> ::= the current center frequency in NR3 format.
The range
of legal values is from -25 GHz to 25 GHz.
See Also
•
"Introduction to :FUNCtion<m> Commands"
•
•
•
NOTE
After a *RST (Reset) or :AUToscale command, the values returned by the
:FUNCtion<m>[:FFT]:CENTer? and :FUNCtion<m>:SPAN? queries depend on the current
:TIMebase:RANGe value. Once you change either the :FUNCtion<m>[:FFT]:CENTer or
:FUNCtion<m>:SPAN value, they no longer track the :TIMebase:RANGe value.
Summary of Contents for P9241
Page 1: ...Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Page 6: ...6 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Page 36: ...36 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 1 What s New...
Page 41: ...Setting Up 2 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 41...
Page 44: ...44 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 2 Setting Up...
Page 222: ...222 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 7 Root Commands...
Page 244: ...244 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands...
Page 256: ...256 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands...
Page 278: ...278 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 10 CHANnel n Commands...
Page 328: ...328 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands...
Page 386: ...386 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands...
Page 390: ...390 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands...
Page 416: ...416 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands...
Page 498: ...498 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 21 MEASure Commands...
Page 532: ...532 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 22 MTESt Commands...
Page 572: ...572 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 24 SAVE Commands...
Page 900: ...900 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 28 TIMebase Commands...
Page 1080: ...1080 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 31 WGEN w Commands...
Page 1142: ...1142 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 34 Error Messages...
Page 1344: ...1344 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide Index...