:MEASure Commands
21
Keysight P9241/42/43A Oscilloscopes SCPI Programmer's Guide
443
:MEASure:DELay
Command Syntax
:MEASure:DELay [<edge_select_mode>][,][<source1>][,<source2>]
<edge_select_mode> ::= {MANual | AUTO}
<source1>, <source2> ::= {CHANnel<n> | FUNCtion<m> | MATH<m> | WMEMory<r
>}
<n> ::= 1 to (# analog channels) in NR1 format
<m> ::= 1 to (# math functions) in NR1 format
<r> ::= 1 to (# ref waveforms) in NR1 format
The :MEASure:DELay command places the instrument in the continuous
measurement mode and starts a delay measurement.
The measurement is taken as:
delay = t(<edge_spec_2>) - t(<edge_spec_1>)
where the <edge_spec> definitions are determined by the <edge_select_mode>
and the :MEASure:DELay:DEFine command.
When the <edge_select_mode> is:
• AUTO — edges are automatically selected: the source1 edge closest to the
timebase reference point is used, and the source2 edge closest to the source1
edge is used.
• MANual — edge numbers are determined by the :MEASure:DELay:DEFine
settings.
Edge numbers greater than zero (0) are counted from the left side of the display
for both sources.
Edge slopes are always determined by the :MEASure:DELay:DEFine settings.
If the <edge_select_mode> is not included with the command, AUTO is the default.
Query Syntax
:MEASure:DELay? [<edge_select_mode>][,][<source1>][,<source2>]
The :MEASure:DELay? query measures and returns the delay between source1 and
source2. Delay measurement slope and edge count parameters are determined by
the <edge_select_mode> and the :MEASure:DELay:DEFine command.
The <edge_select_mode> definitions are the same as with the command syntax.
However, if the <edge_select_mode> is not included with the query, MANual is the
default.
In the :MEASure:DEFine command, you can set upper, middle, and lower threshold
values.
It is the middle threshold value that is used when performing the delay
query.
The standard upper, middle, and lower measurement thresholds are 90%,
Summary of Contents for P9241
Page 1: ...Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Page 6: ...6 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Page 36: ...36 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 1 What s New...
Page 41: ...Setting Up 2 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 41...
Page 44: ...44 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 2 Setting Up...
Page 222: ...222 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 7 Root Commands...
Page 244: ...244 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands...
Page 256: ...256 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands...
Page 278: ...278 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 10 CHANnel n Commands...
Page 328: ...328 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands...
Page 386: ...386 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands...
Page 390: ...390 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands...
Page 416: ...416 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands...
Page 498: ...498 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 21 MEASure Commands...
Page 532: ...532 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 22 MTESt Commands...
Page 572: ...572 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 24 SAVE Commands...
Page 900: ...900 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 28 TIMebase Commands...
Page 1080: ...1080 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 31 WGEN w Commands...
Page 1142: ...1142 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 34 Error Messages...
Page 1344: ...1344 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide Index...