:MEASure Commands
21
Keysight P9241/42/43A Oscilloscopes SCPI Programmer's Guide
439
:MEASure:COUNter
Command Syntax
:MEASure:COUNter [<source>]
<source> ::= {CHANnel<n> | EXTernal}
<n> ::= 1 to (# analog channels) in NR1 format
The :MEASure:COUNter command installs a screen measurement and starts a
counter measurement. If the optional source parameter is specified, the current
source is modified. Any channel except Math or Reference Waveforms may be
selected for the source.
The counter measurement counts trigger level crossings within a certain amount
of time (gate time) and displays the results in Hz.
The gate time is the horizontal range of the oscilloscope but is limited to >= 0.1 s
and <= 10 s. Unlike other measurements, the Zoom horizontal timebase window
does not gate the Counter measurement.
The Counter measurement can measure frequencies up to the bandwidth of the
oscilloscope. The minimum frequency supported is 2.0 / gateTime.
Only one counter measurement may be displayed at a time.
Query Syntax
:MEASure:COUNter? [<source>]
The :MEASure:COUNter? query measures and outputs the counter frequency of
the specified source.
Return Format
<source><NL>
<source> ::= count in Hertz in NR3 format
See Also
•
"Introduction to :MEASure Commands"
•
•
•
NOTE
This command is not available if the source is MATH.
NOTE
The :MEASure:COUNter? query times out if the counter measurement is installed on the front
panel. Use :MEASure:CLEar to remove the front-panel measurement before executing the
:MEASure:COUNter? query.
Summary of Contents for P9241
Page 1: ...Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Page 6: ...6 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Page 36: ...36 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 1 What s New...
Page 41: ...Setting Up 2 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 41...
Page 44: ...44 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 2 Setting Up...
Page 222: ...222 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 7 Root Commands...
Page 244: ...244 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands...
Page 256: ...256 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands...
Page 278: ...278 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 10 CHANnel n Commands...
Page 328: ...328 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands...
Page 386: ...386 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands...
Page 390: ...390 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands...
Page 416: ...416 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands...
Page 498: ...498 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 21 MEASure Commands...
Page 532: ...532 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 22 MTESt Commands...
Page 572: ...572 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 24 SAVE Commands...
Page 900: ...900 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 28 TIMebase Commands...
Page 1080: ...1080 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 31 WGEN w Commands...
Page 1142: ...1142 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 34 Error Messages...
Page 1344: ...1344 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide Index...