More About Oscilloscope Commands
37
Keysight P9241/42/43A Oscilloscopes SCPI Programmer's Guide
1199
In the oscilloscope programmer's documentation, the short form of a command is
indicated by uppercase characters.
Programs written in long form are easily read and are almost self-documenting.
The short form syntax conserves the amount of controller memory needed for
program storage and reduces I/O activity.
Simple Command Headers
Simple command headers contain a single mnemonic. :AUToscale and :DIGitize
are examples of simple command headers typically used in the oscilloscope. The
syntax is:
<program mnemonic><terminator>
Simple command headers must occur at the beginning of a program message; if
not, they must be preceded by a colon.
When program data must be included with the simple command header (for
example, :DIGitize CHANnel1), white space is added to separate the data from the
header. The syntax is:
<program mnemonic><separator><program data><terminator>
Compound Command Headers
Compound command headers are a combination of two or more program
mnemonics. The first mnemonic selects the subsystem, and the second mnemonic
selects the function within that subsystem. The mnemonics within the compound
message are separated by colons. For example, to execute a single function within
a subsystem:
:<subsystem>:<function><separator><program data><terminator>
For example, :CHANnel1:BWLimit ON
Common Command Headers
Common command headers control IEEE 488.2 functions within the instrument
(such as clear status). Their syntax is:
*<command header><terminator>
No space or separator is allowed between the asterisk (*) and the command
header. *CLS is an example of a common command header.
TIMebase
TIM
DELay
DEL
TYPE
TYPE
Long Form
Short form
Summary of Contents for P9241
Page 1: ...Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Page 6: ...6 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Page 36: ...36 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 1 What s New...
Page 41: ...Setting Up 2 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 41...
Page 44: ...44 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 2 Setting Up...
Page 222: ...222 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 7 Root Commands...
Page 244: ...244 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands...
Page 256: ...256 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands...
Page 278: ...278 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 10 CHANnel n Commands...
Page 328: ...328 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands...
Page 386: ...386 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands...
Page 390: ...390 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands...
Page 416: ...416 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands...
Page 498: ...498 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 21 MEASure Commands...
Page 532: ...532 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 22 MTESt Commands...
Page 572: ...572 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 24 SAVE Commands...
Page 900: ...900 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 28 TIMebase Commands...
Page 1080: ...1080 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 31 WGEN w Commands...
Page 1142: ...1142 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 34 Error Messages...
Page 1344: ...1344 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide Index...