242
Keysight P9241/42/43A Oscilloscopes SCPI Programmer's Guide
8
:ACQuire Commands
:ACQuire:TYPE
Command Syntax
:ACQuire:TYPE <type>
<type> ::= {NORMal | AVERage | HRESolution | PEAK}
The :ACQuire:TYPE command selects the type of data acquisition that is to take
place. The acquisition types are:
• NORMal — sets the oscilloscope in the normal mode.
• AVERage — sets the oscilloscope in the averaging mode. You can set the count
by sending the :ACQuire:COUNt command followed by the number of averages.
In this mode, the value for averages is an integer from 1 to 65536. The COUNt
value determines the number of averages that must be acquired.
The AVERage type is not available when in segmented memory mode
(:ACQuire:MODE SEGMented).
• HRESolution — sets the oscilloscope in the high-resolution mode (also known
as
smoothing
). This mode is used to reduce noise at slower sweep speeds
where the digitizer samples faster than needed to fill memory for the displayed
time range.
For example, if the digitizer samples at 200 MSa/s, but the effective sample
rate is 1 MSa/s (because of a slower sweep speed), only 1 out of every 200
samples needs to be stored. Instead of storing one sample (and throwing others
away), the 200 samples are averaged together to provide the value for one
display point. The slower the sweep speed, the greater the number of samples
that are averaged together for each display point.
• PEAK — sets the oscilloscope in the peak detect mode. In this mode,
:ACQuire:COUNt has no meaning.
The AVERage and HRESolution types can give you extra bits of vertical resolution.
See the
User's Guide
for an explanation. When getting waveform data acquired
using the AVERage and HRESolution types, be sure to use the WORD or ASCii
waveform data formats to get the extra bits of vertical resolution.
Query Syntax
:ACQuire:TYPE?
The :ACQuire:TYPE? query returns the current acquisition type.
Return Format
<acq_type><NL>
<acq_type> ::= {NORM | AVER | HRES | PEAK}
NOTE
The obsolete command ACQuire:TYPE:REALtime is functionally equivalent to sending
ACQuire:MODE RTIME; TYPE NORMal.
Summary of Contents for P9241
Page 1: ...Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Page 6: ...6 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Page 36: ...36 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 1 What s New...
Page 41: ...Setting Up 2 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 41...
Page 44: ...44 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 2 Setting Up...
Page 222: ...222 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 7 Root Commands...
Page 244: ...244 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands...
Page 256: ...256 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands...
Page 278: ...278 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 10 CHANnel n Commands...
Page 328: ...328 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands...
Page 386: ...386 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands...
Page 390: ...390 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands...
Page 416: ...416 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands...
Page 498: ...498 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 21 MEASure Commands...
Page 532: ...532 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 22 MTESt Commands...
Page 572: ...572 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 24 SAVE Commands...
Page 900: ...900 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 28 TIMebase Commands...
Page 1080: ...1080 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 31 WGEN w Commands...
Page 1142: ...1142 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 34 Error Messages...
Page 1344: ...1344 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide Index...