Keysight P9241/42/43A Oscilloscopes SCPI Programmer's Guide
30
:WAVeform Commands
:WAVeform:POINts:MODE
Command Syntax
:WAVeform:POINts:MODE <points_mode>
<points_mode> ::= {NORMal | MAXimum | RAW}
The :WAVeform:POINts:MODE command sets the data record to be transferred
with the :WAVeform:DATA? query.
For the analog sources, there are two different records that can be transferred:
• The first is the raw acquisition record. The maximum number of points available
in this record is returned by the :ACQuire:POINts? query. The raw acquisition
record can only be retrieved from the analog sources.
• The second is referred to as the
measurement record
and is a 65,535-point
(maximum) representation of the raw acquisition record. The measurement
record can be retrieved from any source.
If the <points_mode> is NORMal, the measurement record is retrieved.
If the <points_mode> is RAW, the raw acquisition record is used. Under some
conditions, this data record is unavailable.
If the <points_mode> is MAXimum, whichever record contains the maximum
amount of points is used. Usually, this is the raw acquisition record. But, the
measurement record may have more data. If data is being retrieved as the
oscilloscope is stopped and as the data displayed is changing, the data being
retrieved can switch between the measurement and raw acquisition records.
Considerations for
MAXimum or RAW
data retrieval
• The instrument must be stopped (see the :STOP command (see
) or
) in the root subsystem) in order to return
more than the
measurement record
.
• :TIMebase:MODE must be set to MAIN.
• :ACQuire:TYPE must be set to NORMal, AVERage, or HRESolution.
• MAXimum or RAW will allow up to 4,000,000 points to be returned. The
number of points returned will vary as the instrument's configuration is
changed. Use the :WAVeform:POINts? MAXimum query to determine the
maximum number of points that can be retrieved at the current settings.
Query Syntax
:WAVeform:POINts:MODE?
The :WAVeform:POINts:MODE? query returns the current points mode. Setting the
points mode will affect what data is transferred. See the discussion above.
NOTE
If the :WAVeform:SOURce is not an analog or digital source, the only valid parameters for
WAVeform:POINts:MODE is NORMal or MAXimum.
Summary of Contents for P9241
Page 1: ...Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Page 6: ...6 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Page 36: ...36 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 1 What s New...
Page 41: ...Setting Up 2 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 41...
Page 44: ...44 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 2 Setting Up...
Page 222: ...222 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 7 Root Commands...
Page 244: ...244 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands...
Page 256: ...256 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands...
Page 278: ...278 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 10 CHANnel n Commands...
Page 328: ...328 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands...
Page 386: ...386 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands...
Page 390: ...390 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands...
Page 416: ...416 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands...
Page 498: ...498 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 21 MEASure Commands...
Page 532: ...532 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 22 MTESt Commands...
Page 572: ...572 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 24 SAVE Commands...
Page 900: ...900 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 28 TIMebase Commands...
Page 1080: ...1080 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 31 WGEN w Commands...
Page 1142: ...1142 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 34 Error Messages...
Page 1344: ...1344 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide Index...