:FRANalysis Commands
16
Keysight P9241/42/43A Oscilloscopes SCPI Programmer's Guide
343
:FRANalysis:WGEN:VOLTage
Command Syntax
:FRANalysis:WGEN:VOLTage <amplitude>,[<range>]
<amplitude> ::= amplitude in volts in NR3 format
<range> ::= {F20HZ | F100HZ | F1KHZ | F10KHZ | F100KHZ | F1MHZ
| F10MHZ | F20MHZ}
The :FRANalysis:WGEN:VOLTage command specifies the waveform generator's
output sine wave amplitude.
Use the :WGEN:VOLTage:OFFSet command to specify the offset voltage or DC
level.
You can also specify the amplitude and offset indirectly using the
:WGEN:VOLTage:HIGH and :WGEN:VOLTage:LOW commands. For example, an
amplitude of 5 V and an offset of 1 V is the same as a high-level voltage of 4 V and
a low-level voltage of -1 V.
When the amplitude profile setting is on (:FRANalysis:WGEN:VOLTage:PROFile
ON), the <range> option specifies the initial ramp amplitude at the frequency
setting. Amplitudes ramp between the settings specified for individual frequencies.
Query Syntax
:FRANalysis:WGEN:VOLTage? [<range>]
The :FRANalysis:WGEN:VOLTage? query returns the currently specified waveform
generator amplitude.
When the amplitude profile setting is on (:FRANalysis:WGEN:VOLTage:PROFile
ON), the <range> option specifies the frequency whose initial ramp amplitude is
returned.
Return Format
<amplitude><NL>
<amplitude> ::= amplitude in volts in NR3 format
See Also
•
":FRANalysis:WGEN:VOLTage:PROFile"
•
•
•
•
•
•
•
•
•
Summary of Contents for P9241
Page 1: ...Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Page 6: ...6 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Page 36: ...36 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 1 What s New...
Page 41: ...Setting Up 2 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 41...
Page 44: ...44 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 2 Setting Up...
Page 222: ...222 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 7 Root Commands...
Page 244: ...244 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands...
Page 256: ...256 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands...
Page 278: ...278 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 10 CHANnel n Commands...
Page 328: ...328 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands...
Page 386: ...386 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands...
Page 390: ...390 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands...
Page 416: ...416 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands...
Page 498: ...498 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 21 MEASure Commands...
Page 532: ...532 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 22 MTESt Commands...
Page 572: ...572 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 24 SAVE Commands...
Page 900: ...900 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 28 TIMebase Commands...
Page 1080: ...1080 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 31 WGEN w Commands...
Page 1142: ...1142 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 34 Error Messages...
Page 1344: ...1344 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide Index...