:MEASure Commands
:MEASure:FFT:OBW
Command Syntax
:MEASure:FFT:OBW <percentage>[,<source>]
<percentage> ::= percent of spectral power occupied bandwidth is
measured for (in NR3 format)
<source> ::= {FUNCtion<m> | MATH<m> | FFT} (must be an FFT waveform)
<m> ::= 1 to (# math functions) in NR1 format
The :MEASure:FFT:OBW command installs an FFT analysis Occupied Bandwidth
measurement on screen.
Occupied Bandwidth measures the bandwidth (frequency range) containing some
percent (usually 99%) of the total spectral power. While 99% is the industry norm,
you can specify the percent you want to use in the measurement.
The center frequency used in the measurement is the one defined for the FFT
function, and the FFT span represents the total spectral power.
When this measurement is tracked with cursors, the cursors show the measured
bandwidth (frequency range).
Query Syntax
:MEASure:FFT:OBW? <percentage>[,<source>]
The :MEASure:FFT:OBW? query returns the measured Occupied Bandwidth value.
Return Format
<return_value><NL>
<return_value> ::= occupied bandwidth, Hz in NR3 format
See Also
•
•
•
Summary of Contents for P9241
Page 1: ...Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Page 6: ...6 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Page 36: ...36 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 1 What s New...
Page 41: ...Setting Up 2 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 41...
Page 44: ...44 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 2 Setting Up...
Page 222: ...222 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 7 Root Commands...
Page 244: ...244 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands...
Page 256: ...256 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands...
Page 278: ...278 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 10 CHANnel n Commands...
Page 328: ...328 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands...
Page 386: ...386 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands...
Page 390: ...390 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands...
Page 416: ...416 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands...
Page 498: ...498 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 21 MEASure Commands...
Page 532: ...532 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 22 MTESt Commands...
Page 572: ...572 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 24 SAVE Commands...
Page 900: ...900 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 28 TIMebase Commands...
Page 1080: ...1080 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 31 WGEN w Commands...
Page 1142: ...1142 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 34 Error Messages...
Page 1344: ...1344 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide Index...