Keysight B2980B User’s Guide, Edition 1
133
Front Panel Reference
Hard Keys and Rotary Knob
•
TRIGger acquire timer period (Measure Period): 10 ms (this is automatically
extended if the measurement is not completed.)
•
Trigger delay (Source Delay = Measure Delay): 0 s
•
Trigger output: Disable
Function Keys
Six function keys are placed below the LCD. They are assigned to the softkeys,
SPEED, RANGE, Zoom, Clear, Apps, System Menu, and so on. The softkey
assignment depends on the view mode (Meter, Graph, Histogram, or Roll).
Assist Keys
Five assist keys are placed right to the LCD. They are assigned to softkeys which
assist you in changing the measurement setup. The softkey assignment depends
on the view mode and the setup field in the EDIT (green) status.
View
Switches the softkey assignment for the function keys as follows.
•
Softkeys available for the present view mode
•
Softkeys for selecting the view mode
The following softkeys are available. Press one for selecting the view mode.
Meter View
Displays the Meter view.
Graph View
Displays the Graph view.
Histogram View
Displays the Histogram view.
Roll View
Displays the Roll view.
Cancel / Local
Cancels the setup operation if the instrument is in the local status. Returns the
instrument to the local status if it is in the remote status.
Measurement
navigation keys
Changes the measurement speed (SPEED) and measurement range (RANGE)
settings. The following keys are available.
Coarse Res
Increases measurement speed (decreases the aperture time).
The setting value is displayed on the SPEED function key.
In the auto aperture mode, changes the aperture time in
“Stable
Normal
Quick” order.
In the manual aperture mode, changes the aperture time in
“100.0 PLC
10.0 PLC
1.0 PLC
0.1 PLC
0.01 PLC
0.001 PLC” order.
Summary of Contents for B2980B Series
Page 13: ...Keysight B2985B B2987B High Low 1050 Vdc 3 Interlock High Low On Off On Off HV Interlock...
Page 16: ......
Page 24: ......
Page 52: ...52 Keysight B2980B User s Guide Edition 1 Introduction Options...
Page 80: ...80 Keysight B2980B User s Guide Edition 1 Installation Using Digital I O and Trigger In Out...
Page 202: ...202 Keysight B2980B User s Guide Edition 1 Front Panel Reference System Key Group...
Page 264: ...264 Keysight B2980B User s Guide Edition 1 Function Details Initial Settings...