Keysight B2980B User’s Guide, Edition 1
229
Function Details
Limit Test
Limit Test
Limit test is a pass/fail judgment performed for a measurement data or math
result data obtained by a channel. It can be performed if both individual limit test
and composite limit test are set to ON. Maximum of twelve limit tests can be
defined and used for the bins of composite limit test.
Composite limit test supports the following two operation modes.
•
Grading mode
Performs limit test for up to 12 test limits (bins) until a failure is detected. See
for an example flowchart.
•
Sorting mode
Performs limit test for up to 12 test limits (bins) until a pass is detected. See
for an example flowchart.
It also supports the following communication functionality via the Digital I/O
connector and can classify test devices using the component handler.
•
Receiving the SOT (start of test) strobe pulse
•
Transmitting the EOT (end of test) strobe pulse
•
Transmitting the BUSY signal
•
Transmitting the bit pattern to show a pass/fail judgement result
In the figures, SOT is the start-of-test strobe pulse sent by the component handler
connected to B2980 via the Digital I/O connector.
To set the composite limit test, see
“Composite Limit Test Setup Dialog Box” on
To set the individual limit tests, see
“Limit Test Setup Dialog Box” on page 180
The composite limit test result Pass or Fail is displayed with the measurement
result data on the Meter view. To display the limit test result log, see
Summary of Contents for B2980B Series
Page 13: ...Keysight B2985B B2987B High Low 1050 Vdc 3 Interlock High Low On Off On Off HV Interlock...
Page 16: ......
Page 24: ......
Page 52: ...52 Keysight B2980B User s Guide Edition 1 Introduction Options...
Page 80: ...80 Keysight B2980B User s Guide Edition 1 Installation Using Digital I O and Trigger In Out...
Page 202: ...202 Keysight B2980B User s Guide Edition 1 Front Panel Reference System Key Group...
Page 264: ...264 Keysight B2980B User s Guide Edition 1 Function Details Initial Settings...