148
Keysight B2980B User’s Guide, Edition 1
Front Panel Reference
Display and Assist/Function Keys
4. After that, repeats measuring the DUT as many times as is defined by the
Repeat Count with a defined Interval time between measurements. The
measurement data is displayed on the front display after each measurement,
and all measurement data is stored in the trace buffer.
5. After all measurements are completed, turns the applied voltage off (0 V) and
turns the input/output switches off.
The sequential measurement is appropriate for measuring time-related
characteristics, and can perform a repeating measurement with simple setting. It
is useful for volume and sheet resistivity measurements.
To prevent electrical shock, discharge the DUT sufficiently before you touch the
DUT after measurement.
It is useful to set the output-off state to ZERO for discharging before and after
the measurement.
To perform the sequential measurement, do the steps below.
1. Complete the following measurement setup.
•
Measurement type (current/charge/voltage/resistance)
•
Measurement speed (aperture mode and time)
•
Measurement range: FIXED
•
Limit test, if necessary
•
Handler control, if necessary
2. Press the Apps function key on the Meter view, then press the #3 assist key.
The Sequential Measurement dialog box opens.
3. Set the following parameters in this dialog box.
Source Value
Forcing voltage.
Charge Time
Wait time after starting the voltage forcing until performing
the 1st measurement.
Interval
Interval between measurements.
Repeat Count
Repeat count of measurements.
4. Press OK (function key) to start the sequential measurement.
Summary of Contents for B2980B Series
Page 13: ...Keysight B2985B B2987B High Low 1050 Vdc 3 Interlock High Low On Off On Off HV Interlock...
Page 16: ......
Page 24: ......
Page 52: ...52 Keysight B2980B User s Guide Edition 1 Introduction Options...
Page 80: ...80 Keysight B2980B User s Guide Edition 1 Installation Using Digital I O and Trigger In Out...
Page 202: ...202 Keysight B2980B User s Guide Edition 1 Front Panel Reference System Key Group...
Page 264: ...264 Keysight B2980B User s Guide Edition 1 Function Details Initial Settings...