How to Perform Measurement
Charge Measurement
Figure 3-17
Charge Measurement Connection
Procedure
You can perform the charge measurement as follows. Also see
.
Step 1. Set the automatic discharge function. See
“Automatic Discharge” on page 122
.
Step 2. Press the
View
key and select the Meter View function key to display the Meter
view.
Step 3. Press the
COUL. (Q)
assist key to set the charge measurement mode.
Step 4. Use the RANGE function key and the navigation keys to set the measurement
range you want to use. See
for more
information.
Step 5. Use the SPEED function key and the navigation keys to set the measurement
speed (aperture time) you desire. See
for more
information.
Step 6. Press the
Filter
key to open the Measure Filter dialog box. And set the
measurement filter on the dialog box. See
for more
information.
Step 7. Connect measurement charge (DUT) to the Ammeter input connector. See
for the connection.
Step 8. Press the Ammeter
On/Off
switch to enable the coulomb meter. This turns the
switch green.
Step 9. Start measurement as follows.
Qs
Shield
Input
Chassis ground
Triaxial cable
Banana-to-lug cable
Common is connected to chassis ground by using Banana-to-lug cable.
Triaxial bulkhead connector
Qs
A
Summary of Contents for B2980B Series
Page 13: ...Keysight B2985B B2987B High Low 1050 Vdc 3 Interlock High Low On Off On Off HV Interlock...
Page 16: ......
Page 24: ......
Page 52: ...52 Keysight B2980B User s Guide Edition 1 Introduction Options...
Page 80: ...80 Keysight B2980B User s Guide Edition 1 Installation Using Digital I O and Trigger In Out...
Page 202: ...202 Keysight B2980B User s Guide Edition 1 Front Panel Reference System Key Group...
Page 264: ...264 Keysight B2980B User s Guide Edition 1 Function Details Initial Settings...