218
Keysight B2980B User’s Guide, Edition 1
Function Details
Voltage Source (B2985B/B2987B)
List Sweep Output
The list sweep function is effective for performing an arbitrary waveform output.
For example, the voltage source can apply the waveform as shown in
and the B2985B/B2987B can perform the measurement at each output value. The
source output and the measurement can be performed in the minimum interval
100
s.
The voltage source output and measurement timing is controlled by the trigger
system. And the interval can be set to a constant value if the trigger type is set to
TIMER.
To set the trigger system, see
“Trigger Setup Sub-panel” on page 157
. The trigger
interval is set by the Period parameter. The number of output values is set by the
Count parameter.
To set the voltage source output values, use the List Sweep dialog box. See
shows an example setup of the List Sweep dialog box with an image of
the output waveform.
To set the voltage output range, use the Spot Source Range field, see
Figure 5-5
List Sweep Dialog Box
Summary of Contents for B2980B Series
Page 13: ...Keysight B2985B B2987B High Low 1050 Vdc 3 Interlock High Low On Off On Off HV Interlock...
Page 16: ......
Page 24: ......
Page 52: ...52 Keysight B2980B User s Guide Edition 1 Introduction Options...
Page 80: ...80 Keysight B2980B User s Guide Edition 1 Installation Using Digital I O and Trigger In Out...
Page 202: ...202 Keysight B2980B User s Guide Edition 1 Front Panel Reference System Key Group...
Page 264: ...264 Keysight B2980B User s Guide Edition 1 Function Details Initial Settings...