Keysight B2980B User’s Guide, Edition 1
101
How to Perform Measurement
Voltage Measurement
Figure 3-10
Guarding
Never connect the Guard terminal to any output, including circuit common,
chassis ground, or any other guard terminal. Doing so will damage the
instrument.
V
Buffer amplifier
Guard
Voltage
under test
Common
Output
resistance
V
Insulation
resistance
Voltage
under test
Common
Output
resistance
Cable
capacitance
Summary of Contents for B2980B Series
Page 13: ...Keysight B2985B B2987B High Low 1050 Vdc 3 Interlock High Low On Off On Off HV Interlock...
Page 16: ......
Page 24: ......
Page 52: ...52 Keysight B2980B User s Guide Edition 1 Introduction Options...
Page 80: ...80 Keysight B2980B User s Guide Edition 1 Installation Using Digital I O and Trigger In Out...
Page 202: ...202 Keysight B2980B User s Guide Edition 1 Front Panel Reference System Key Group...
Page 264: ...264 Keysight B2980B User s Guide Edition 1 Function Details Initial Settings...