212
Keysight B2980B User’s Guide, Edition 1
Function Details
Null, Offset Cancel
Resistance Measurement (B2985B/B2987B)
If the Null key is pressed when the NULL indicator is turned off, the current
resistance measurement value is set as the offset value. Just after that, the
resistance measurement is performed once, and the value that subtracted the
offset value from the actual measured value is displayed as the measurement
result. And also the NULL indicator is turned on.
If the Null key is pressed when the NULL indicator is tuned on, the null function is
disabled and the NULL indicator is turned off. Just after that, the resistance
measurement is performed once, and the actual measurement value is displayed.
Summary of Contents for B2980B Series
Page 13: ...Keysight B2985B B2987B High Low 1050 Vdc 3 Interlock High Low On Off On Off HV Interlock...
Page 16: ......
Page 24: ......
Page 52: ...52 Keysight B2980B User s Guide Edition 1 Introduction Options...
Page 80: ...80 Keysight B2980B User s Guide Edition 1 Installation Using Digital I O and Trigger In Out...
Page 202: ...202 Keysight B2980B User s Guide Edition 1 Front Panel Reference System Key Group...
Page 264: ...264 Keysight B2980B User s Guide Edition 1 Function Details Initial Settings...