Keysight B2980B User’s Guide, Edition 1
115
How to Perform Measurement
Resistance Measurement
To specify the low terminal status, perform the following procedure. Setting is not
saved. Turning off the instrument sets the status to CCOM. This procedure can be
performed only if the voltage source output is Off.
1. Press the System Menu > Config > Source > Connection function keys. This
opens the Output Connection dialog box.
2. Set the Low Terminal State on this dialog box. It must be FLOATING for making
the floating condition, or CCOM for connecting to the circuit common.
3. Press OK.
For the floating status, the Floating indicator appears on the
status information area.
Although the Low terminal is set to FLOATING, the High potential and the Low
potential must be less than or equal to
1050 V for chassis ground. The output
voltage is limited to
1050 V even if the Voltage Source makes the cascade
connection with an external voltage source.
Do not apply current to the chassis ground. Doing so will damage the
instrument.
DC Voltage Output Range
Voltage Source has three output ranges, 20 V range,
1000 V range, and
1000 V
range, for applying DC voltage.
To set the output range, press the
20 V
,
+1000 V
, or
-1000 V
assist key available
when the Spot Source Range field is in the EDIT status. This field is in the range
setup sub-panel on the Meter view.
To display the sub-panel, you may need to press the
More... 1 of 3
assist key and
the
Hide Hist.
,
Hide Roll
, or
Hide Trigger
assist key.
Staircase Sweep Output Setup
Voltage Source supports the following staircase sweep output.
•
Linear single sweep
Voltage Source applies staircase sweep voltage from start to stop in a step.
This sweep is available when VS Function is set to LINEAR SINGLE.
•
Linear double sweep
Summary of Contents for B2980B Series
Page 13: ...Keysight B2985B B2987B High Low 1050 Vdc 3 Interlock High Low On Off On Off HV Interlock...
Page 16: ......
Page 24: ......
Page 52: ...52 Keysight B2980B User s Guide Edition 1 Introduction Options...
Page 80: ...80 Keysight B2980B User s Guide Edition 1 Installation Using Digital I O and Trigger In Out...
Page 202: ...202 Keysight B2980B User s Guide Edition 1 Front Panel Reference System Key Group...
Page 264: ...264 Keysight B2980B User s Guide Edition 1 Function Details Initial Settings...