210
Keysight B2980B User’s Guide, Edition 1
Function Details
Null, Offset Cancel
Null, Offset Cancel
The Null (offset cancel) function displays the measurement result based on setting
the measured value under specific condition as reference. The value measured on
pressing the Null key is stored as offset value, and then the value that subtracts
the offset value from the actual measurement value is displayed as the
measurement value. If this function is enabled, the NULL indicator turns on. If the
NULL indicator turns off, the actual measurement value is displayed.
If the measurement mode is changed when the NULL indicator turns on, the NULL
indicator turns off, even though the null function remains enabled. This state will
be kept until the null function for the primary measurement mode is enabled.
The following sections show the usage examples of this function:
Current Measurement (B2981B/B2983B/B2985B/B2987B)
and Charge Measurement (B2985B/B2987B)
1. Prepare the reference connection for measurement.
This function is used to reject the offset created from the measurement
environment. So, connect the measurement cables to the instrument, and then
make the measurement terminals open immediately close to the device under
test (DUT).
Or, you can use this function by using the state where DUT is connected as
reference state.
2. Turn the Ammeter On.
3. Press the
Run/Stop
key to perform repeat measurement.
If the measured value is not stable, adjust the measurement speed (aperture
time) and the measurement filter setting.
4. Confirm that the measurement value is stable, and then press the Null key.
The NULL indicator turns on, and the null function is enabled.
To disable this function, press the Null key during Ammeter On, then the NULL
indicator turns off.
Summary of Contents for B2980B Series
Page 13: ...Keysight B2985B B2987B High Low 1050 Vdc 3 Interlock High Low On Off On Off HV Interlock...
Page 16: ......
Page 24: ......
Page 52: ...52 Keysight B2980B User s Guide Edition 1 Introduction Options...
Page 80: ...80 Keysight B2980B User s Guide Edition 1 Installation Using Digital I O and Trigger In Out...
Page 202: ...202 Keysight B2980B User s Guide Edition 1 Front Panel Reference System Key Group...
Page 264: ...264 Keysight B2980B User s Guide Edition 1 Function Details Initial Settings...