Keysight B2980B User’s Guide, Edition 1
Front Panel Reference
Config Key Group
Auto Dis.
Automatic discharge function setup, ON or
OFF.
The softkey displays the present ON/OFF
status of this function. Pressing this key
switches the status.
Dis. Level
Minimum discharge level, 2 nC, 20 nC,
200 nC, or 2
C.
The softkey displays the present discharge
level. Pressing this key switches the level.
Ohms
(B2985B/B2987B) Displays the following softkeys for setting
the resistance measurement operation.
V Select
Resistance calculation mode setup. Source
or Measure.
If Source is selected, the resistance is
calculated by R = Vs/Im. Where, Vs is the
voltage source output value, and Im is the
current measurement value.
If Measure is selected, the resistance is
calculated by R = Vm/Im. Where, Vm is the
voltage measurement value, and Im is the
current measurement value.
The softkey displays the present
Source/Measure status of this function.
Pressing this key switches the status.
V Control
V Control mode setup. Manual or Auto. For
details, see
If Manual is set, the resistance measurement
is performed with the condition given by the
Measure Amps, Spot Source Range, and
Voltage Source field settings.
If Auto is set, the resistance measurement is
performed with the condition automatically
derived from the Measure Ohms field
setting.
Summary of Contents for B2980B Series
Page 13: ...Keysight B2985B B2987B High Low 1050 Vdc 3 Interlock High Low On Off On Off HV Interlock...
Page 16: ......
Page 24: ......
Page 52: ...52 Keysight B2980B User s Guide Edition 1 Introduction Options...
Page 80: ...80 Keysight B2980B User s Guide Edition 1 Installation Using Digital I O and Trigger In Out...
Page 202: ...202 Keysight B2980B User s Guide Edition 1 Front Panel Reference System Key Group...
Page 264: ...264 Keysight B2980B User s Guide Edition 1 Function Details Initial Settings...