168
Keysight B2980B User’s Guide, Edition 1
Front Panel Reference
Display and Assist/Function Keys
Status Information
The status information is common for all display modes. It is displayed above the
bottom softkeys associated with the function keys.
Table 4-6
Status Indicator
1
Label
Color
Description
AUTO
white
Trigger auto. Automatic trigger is enabled.
ARM
white
Trigger active. Trigger system is active.
HV
yellow
High voltage. Output voltage setting is over
21 V.
The low terminal of the voltage source is disconnected from the circuit common, and is in
floating status.
The inner shield of the Voltmeter input connector (triaxial) is connected to the guard.
Battery operation indicator (B2983B/B2987B). This indicates the remaining battery level
when the instrument is powered by the battery.
REM
white
Remote. Instrument is in the remote state.
Local lockout.
Instrument is in the local lockout (LLO) status.
ERR
white
Error. At least one error is detected.
EDIT
green
Edit mode. The field specified by the pointer can be edited. Not in the move mode.
MOVE
blue
Move mode. The field pointer can be moved. Not in the edit mode.
LAN
green or
red
LXI LAN status indicator. Green indicates normal LAN status. Red indicates abnormal
condition. Blinking indicates the LAN identification status.
1
white
Display mode: Meter view
G
white
Display mode: Graph view
R
white
Display mode: Roll view
H
white
Display mode: Histogram view
Summary of Contents for B2980B Series
Page 13: ...Keysight B2985B B2987B High Low 1050 Vdc 3 Interlock High Low On Off On Off HV Interlock...
Page 16: ......
Page 24: ......
Page 52: ...52 Keysight B2980B User s Guide Edition 1 Introduction Options...
Page 80: ...80 Keysight B2980B User s Guide Edition 1 Installation Using Digital I O and Trigger In Out...
Page 202: ...202 Keysight B2980B User s Guide Edition 1 Front Panel Reference System Key Group...
Page 264: ...264 Keysight B2980B User s Guide Edition 1 Function Details Initial Settings...