Keysight B2980B User’s Guide, Edition 1
85
How to Perform Measurement
Current Measurement
Current Measurement
The B2981B/B2983B/B2985B/B2987B supports the current measurement
capability shown in
Table 3-1
Current Measurement Range, Value, and Resolution
Requirements
Before turning the instrument on, connect cable, test leads, test fixture, and so on,
used for the measurement. See
and
for connection examples.
The following accessories can be used.
•
Triaxial cable, 200 V, 1.5 m
•
Triaxial bulkhead connector, if needed
•
Banana to lug cable, for connecting Common to chassis ground
Instead of the triaxial cable and the triaxial bulkhead connector, Keysight N1415A
Triaxial to alligator clip cable, 200 V, 1.5 m can be used.
Range value
Measurement value
Display resolution
2 pA
0
|I|
2.1 pA
1 aA
20 pA
0
|I|
21 pA
10 aA
200 pA
0
|I|
210 pA
100 aA
2 nA
0
|I|
2.1 nA
1 fA
20 nA
0
|I|
21 nA
10 fA
200 nA
0
|I|
210 nA
100 fA
2
A
0
|I|
2.1
A
1 pA
20
A
0
|I|
21
A
10 pA
200
A
0
|I|
210
A
100 pA
2 mA
0
|I|
2.1 mA
1 nA
20 mA
0
|I|
21 mA
10 nA
Summary of Contents for B2980B Series
Page 13: ...Keysight B2985B B2987B High Low 1050 Vdc 3 Interlock High Low On Off On Off HV Interlock...
Page 16: ......
Page 24: ......
Page 52: ...52 Keysight B2980B User s Guide Edition 1 Introduction Options...
Page 80: ...80 Keysight B2980B User s Guide Edition 1 Installation Using Digital I O and Trigger In Out...
Page 202: ...202 Keysight B2980B User s Guide Edition 1 Front Panel Reference System Key Group...
Page 264: ...264 Keysight B2980B User s Guide Edition 1 Function Details Initial Settings...