116
Keysight B2980B User’s Guide, Edition 1
How to Perform Measurement
Resistance Measurement
Voltage Source applies staircase sweep voltage from start to stop in a step and
from stop to start in the same step. This sweep is available when VS Function is
set to LINEAR DOUBLE.
Sweep output is set by the following parameters. They are shown in the source
function sub-panel on the Meter view. To display the sub-panel, press the
Show
VS Func.
assist key. To see the assist key, you may need to press the
More... 1 of
3
assist key and/or the
More... 2 of 3
assist key.
VS Function
LINEAR SINGLE or LINEAR DOUBLE
Start
Sweep start voltage, in V
Stop
Sweep stop voltage, in V
Points
Number of sweep steps between Start and Stop
Step
Sweep step voltage, in V
For Point and Step, specify one. Other one is automatically calculated by the
formula, Point = |Stop
Start|/|Step|
1.
To set the parameter value, use the assist key and/or the rotary knob.
To confirm the sweep output shape, press the
Show Preview
assist key.
Additionally the following setup parameters are available. They are shown in the
Sweep dialog box opened by pressing the System Menu > Config > Source >
Sweep function keys.
Sweep Ranging
Ranging mode FIXED or BEST (automatically uses the
minimum range which covers the whole sweep output)
Sweep Direction
Direction UP (Start to Stop) or DOWN (Stop to Start)
Output after Sweep
Output voltage after sweep, START VALUE or END VALUE
To set the parameter value, use the assist key and/or the rotary knob, and press
OK to apply the settings and close the dialog box.
List Sweep Output Setup
Voltage Source can apply the voltage specified by the list you define.
To set the list sweep output, display the source function sub-panel on the Meter
view by pressing the
Show VS Func.
assist key. To see the assist key, you may
need to press the
More... 1 of 3
assist key and/or the
More... 2 of 3
assist key.
And set LIST to the VS Function field. The sub-panel shows the following
information.
Summary of Contents for B2980B Series
Page 13: ...Keysight B2985B B2987B High Low 1050 Vdc 3 Interlock High Low On Off On Off HV Interlock...
Page 16: ......
Page 24: ......
Page 52: ...52 Keysight B2980B User s Guide Edition 1 Introduction Options...
Page 80: ...80 Keysight B2980B User s Guide Edition 1 Installation Using Digital I O and Trigger In Out...
Page 202: ...202 Keysight B2980B User s Guide Edition 1 Front Panel Reference System Key Group...
Page 264: ...264 Keysight B2980B User s Guide Edition 1 Function Details Initial Settings...