158
Keysight B2980B User’s Guide, Edition 1
Front Panel Reference
Display and Assist/Function Keys
Delay
Sets the trigger delay time.
Period
Sets the trigger period.
Trigger
Selects the trigger source by using the assist key AUTO, BUS,
TIMER, INT
m
, LAN, EXT
n
, or TIN, where
m
is the integer 1 or 2,
and
n
is an integer from between 1 and 7 See
for the
trigger sources.
When the field pointer is in the EDIT (green) status on the Delay or Period field, the
assist keys are changed to the unit keys as shown below.
•
s, ms, s, and ks
Table 4-2
Trigger Types and Setup Parameters
Trigger
= AUTO automatically selects the trigger source best suited for the present
operating mode by using internal algorithms.
Trigger
= BUS uses the remote interface trigger command such as GET, TRIGger,
and *TRG.
Trigger
= TIMER uses a signal internally generated every interval set by the
Period
parameter.
Trigger
= INT1 or INT2 uses a signal from the internal bus 1 or 2, respectively.
Trigger
= LAN uses the LXI trigger.
Trigger
= EXT
n
uses a signal from the DIO pin
n
which is an output port of the
Digital I/O D-sub connector on the rear panel.
n
=1 to 7.
Trigger
= TIN uses a signal from the Trigger In connector on the rear panel.
Type
Count
Delay
Period
Trigger
AUTO
set
automatically
0 second
N.A.
AUTO
SYNC
entered value
entered value
N.A.
AUTO
TIMER
entered value
entered value
entered value
TIMER
MANUAL
entered value
entered value
entered value
selected value
Summary of Contents for B2980B Series
Page 13: ...Keysight B2985B B2987B High Low 1050 Vdc 3 Interlock High Low On Off On Off HV Interlock...
Page 16: ......
Page 24: ......
Page 52: ...52 Keysight B2980B User s Guide Edition 1 Introduction Options...
Page 80: ...80 Keysight B2980B User s Guide Edition 1 Installation Using Digital I O and Trigger In Out...
Page 202: ...202 Keysight B2980B User s Guide Edition 1 Front Panel Reference System Key Group...
Page 264: ...264 Keysight B2980B User s Guide Edition 1 Function Details Initial Settings...