Function Details
Analog Output
Use the fixed measurement range when you use the Analog Out terminal.
For example, if the input current is
20 pA when the measurement range is 20 pA,
the output voltage at the Analog Out terminal is
2.04 V
0.204 V.
You can specify the measurement mode that is applied to the analog output
function using the Analog Signal Output field in the Input Connection dialog box
opened by System Menu > Config > Measure > Connection operations. See
Connection Dialog Box” on page 174
.
shows an example of the connections to a recording device such as
chart recorder using the Analog Out terminal.
The output impedance of the Analog Out terminal is 1 k
. To minimize the effects
of loading, connect a measurement device that the input impedance should be as
high as possible. For example, If a device with an input impedance of 10 M
is
connected, the error due to loading will be approximately 100 ppm.
Figure 5-3
Connections of Analog Out Terminal
Do not connect the Analog Out terminal to earth while floating the input. It may
damage the instrument.
Test leads
Recording Device
High
Low
Banana-to-lug cable
Summary of Contents for B2980B Series
Page 13: ...Keysight B2985B B2987B High Low 1050 Vdc 3 Interlock High Low On Off On Off HV Interlock...
Page 16: ......
Page 24: ......
Page 52: ...52 Keysight B2980B User s Guide Edition 1 Introduction Options...
Page 80: ...80 Keysight B2980B User s Guide Edition 1 Installation Using Digital I O and Trigger In Out...
Page 202: ...202 Keysight B2980B User s Guide Edition 1 Front Panel Reference System Key Group...
Page 264: ...264 Keysight B2980B User s Guide Edition 1 Function Details Initial Settings...