background image

 

E Instruments 

AQ Incuba Air Quality Monitor 

Quick Reference Operating Instructions 

 

NOTE: FIND FULL DETAILED MANUAL ON INCLUDED THUMBDRIVE 

 

NOTE: DO NOT LIFT OR HANDLE THE INSTRUMENT BY THE BLACK  

EXTERNAL TEMP, %RH, PRESSURE SENSOR HOUSING! 

 

 

1.

 

Push the On/Off power button to activate the monitor & sampling pump. Before doing any monitoring allow the unit to run for 

approximately 40-60 seconds for the CO

sensor to stabilize & calibrate. 

 

2.

 

If needed, connect the external black Temperature and Humidity module (Part# TRH Mod) as shown in Figure 1 above. 

 

3.

 

Select a Monitoring Position:  

 

a.

 

If the incubator has a glass/see-thru door and the AQ Incuba can fit safely inside the incubator without 

damaging the instrument (ex: extremely high humidity), place the AQ Incuba with the TRH Module connected 
inside the incubator. If this is not feasible or the instrument needs to remain outside the incubator, please see 

part b. 

 

b.

 

If the instrument needs to remain outside the incubator, connect the Filter/hose assembly (AQ650078) shown 

in Figure 2 to the Center “Sample-In” / “A” inlet at the bottom of the AQ Incuba. If temperature and humidity 
need to be measured connect the optional TRH-EXT probe shown in Figure 3 to the bottom of the AQ Incuba. 

The proper connections are shown in Figure 4 both. Figure 5 shows proper usage of the TRH-EXT and 

AQ650078 together. 

 

 
 

 

 

 

 
 
 
 

 

 

 

 
 

 

 

 

 

 
 

 

 

 

 
 

 

 

Figure 2  

AQ650078 (Included) 

Figure 3 
TRH-EXT (Optional) 

Figure 4 

Figure 1 

Figure 5 

www.

.com

[email protected]

1.800.561.8187

Reviews: