1-8
Performance Verification
Model 2701 Service Manual
Model 2701 verification
Perform these tests to verify accuracy using the Model 2701 front panel terminals.
Verifying DC voltage
Check DC voltage accuracy by applying accurate voltages from the DC voltage calibrator to
the Model 2701 INPUT jacks and verifying that the displayed readings fall within specified
limits.
CAUTION
Do not exceed 1000V peak between front terminals INPUT HI and
INPUT LO because instrument damage may occur.
Follow these steps to verify DC voltage accuracy:
1.
Connect the Model 2701 HI and LO INPUT jacks to the DC voltage calibrator as
shown in
Figure 1-1
. Make sure the INPUTS switch is set to the FRONT position.
NOTE
Use shielded, low-thermal connections when testing the 100mV and 1V ranges to
avoid errors caused by noise or thermal effects. Connect the shield to the
calibrator’s output LO terminal.
Figure 1-1
Connections for Model 2701 DC volts verification
OUTPUT
HI
OUTPUT
LO
INPUT HI
Calibrator (Output DC Voltage)
Note: Use shielded, low-thermal cables
for 100mV and 1V ranges.
Model 2701
INPUT LO
!
Model 2701 Ethernet Multimeter / Data Acquisition System
RANGE
F
500V
PEAK
FRONT/REAR
3A 250V
AMPS
HI
INPUT
LO
SENSE
Ω
4 WIRE
INPUTS
350V
PEAK
1000V
PEAK
AUTO
SHIFT
LOCAL
POWER
RANGE
R
EXIT
ENTER
DIGITS RATE
REL
FILTER
TRIG
EX TRIG
STORE RECALL
OPEN
DCV
DCI
MATH
O U T P U T
RATIO
ACV
ACI
Ω
2
Ω
4
FREQ
TEMP
CH AVG
CONT
PERIOD SENSOR
LIMITS
ON/OFF
DELAY
HOLD
SAVE
SETUP
CONFIG
HALT
TYPE
LSYNC
TEST
MONITOR
STEP
SCAN
OCOMP
CH-OFF
CARD
CLOSE
Integra Series
RS-232
ETHERNET
CAT I
Summary of Contents for 2701
Page 15: ...1 PerformanceVerification...
Page 47: ...2 Calibration...
Page 70: ...3 RoutineMaintenance...
Page 78: ...4 Troubleshooting...
Page 97: ...5 Disassembly...
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ...6 ReplaceableParts...
Page 125: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ...A Specifications...
Page 140: ...B CalibrationReference...