Model 2701 Service Manual
Disassembly
5-7
Instrument reassembly
Reassemble the instrument by reversing the previous disassembly procedures. Make sure
that all parts are properly seated and secured, and that all connections are properly made. To
ensure proper operation, replace and securely fasten the shield.
WARNING
To ensure continued protection against electrical shock, verify that power
line ground (green and yellow wire attached to the power module) and the
power transformer ground (black wire) are connected to the chassis. When
installing the power transformer, be sure to reconnect the black ground
wire to the mounting stud on side of the chassis. Be sure to install the bot-
tom case screws to assure a good case-to-chassis ground connection.
Input terminal wire connections
During reassembly, use the information in
Table 5-1
to connect input terminal wires.
Power module wire connections
Use the information in
Table 5-2
and DETAIL B of drawing 2701-050 to connect power
module wires.
Table 5-1
Input terminal wire colors
Input terminal
Wire color
INPUT HI
INPUT LO
SENSE HI
SENSE LO
AMPS
Red
Black
Yellow
Gray
White
Table 5-2
Power module wire colors
Location
Wire color
Right side
Right top
Left top
Right bottom
Left bottom
Gray
Violet
White
Red
Blue
Summary of Contents for 2701
Page 15: ...1 PerformanceVerification...
Page 47: ...2 Calibration...
Page 70: ...3 RoutineMaintenance...
Page 78: ...4 Troubleshooting...
Page 97: ...5 Disassembly...
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ...6 ReplaceableParts...
Page 125: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ...A Specifications...
Page 140: ...B CalibrationReference...