Model 2701 Service Manual
Performance Verification
1-13
Verifying AC current
Check AC current accuracy by applying accurate AC voltage current at specific frequencies
from the AC current calibrator to the Model 2701 input, verifying that the displayed readings
fall within specified limits. Follow these steps to verify AC current.
1.
Connect the Model 2701 AMPS and INPUT LO jacks to the calibrator as shown in
Figure 1-4
. Be sure the INPUTS switch is in the FRONT position.
Figure 1-4
Connections for Model 2701 AC current verification
2.
Select the AC current function by pressing the
ACI
key.
3.
Set the Model 2701 for the 1A range.
4.
Source 1A and 3A, 1kHz full-scale AC currents as summarized in
Table 1-5
, and verify
that the readings are within stated limits.
Table 1-5
ACI limits
ACV range Applied AC voltage
Reading limits @ 1kHz (1 year, 18°C to 28°C)
1A
3A
1.000000A
3.00000A*
0.99860 to 1.00140A
2.9817 to 3.0183A
* If the Fluke 5725A amplifier is not available, apply 2.2A from the calibrator. Reading limits for 2.2A are
2.1949 to 2.2051A.
!
Model 2701 Ethernet Multimeter / Data Acquisition System
RANGE
F
500V
PEAK
FRONT/REAR
3A 250V
AMPS
HI
INPUT
LO
SENSE
Ω
4 WIRE
INPUTS
350V
PEAK
1000V
PEAK
AUTO
SHIFT
LOCAL
POWER
RANGE
R
EXIT
ENTER
DIGITS RATE
REL
FILTER
TRIG
EX TRIG
STORE RECALL
OPEN
DCV
DCI
MATH
O U T P U T
RATIO
ACV
ACI
Ω
2
Ω
4
FREQ
TEMP
CH AVG
CONT
PERIOD SENSOR
LIMITS
ON/OFF
DELAY
HOLD
SAVE
SETUP
CONFIG
HALT
TYPE
LSYNC
TEST
MONITOR
STEP
SCAN
OCOMP
CH-OFF
CARD
CLOSE
Integra Series
RS-232
ETHERNET
AMPS
Calibrator (Output AC Current)
Model 2701
INPUT LO
OUTPUT
HI
OUTPUT LO
CAT I
Summary of Contents for 2701
Page 15: ...1 PerformanceVerification...
Page 47: ...2 Calibration...
Page 70: ...3 RoutineMaintenance...
Page 78: ...4 Troubleshooting...
Page 97: ...5 Disassembly...
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ...6 ReplaceableParts...
Page 125: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ...A Specifications...
Page 140: ...B CalibrationReference...