1-32
Performance Verification
Model 2701 Service Manual
Verifying ratio and average
Follow the procedure below to verify ratio and average.
CAUTION
Exceeding 300V between plug-in module INPUT or SENSE H and L
terminals may cause instrument damage.
1.
Connect the Model 7700 CH1 and CH11 H and L terminals to the DC calibrator, as
shown in
Figure 1-16
.
2.
Install the Model 7700 in Slot 1 of the Model 2701, then turn on the power and allow
the unit to warm up for two hours before proceeding. Be sure the front panel INPUTS
switch is set to the REAR position.
3.
Select the Model 2701 DCV function and the 1V range. Close Channel 1 by pressing
the CLOSE key and keying in 101.
4.
Select the Model 2701 RATIO function (press SHIFT then RATIO).
5.
Set the calibrator output to 1.00000V DC and allow the reading to settle.
6.
Verify that the ratio reading is between 0.9999926 and 1.000074.
7.
Press OPEN to open Channel 1.
Figure 1-16
Connections for Model 7700 ratio and average verification
H L H L
AMPS
H L H L
H L H L
H L H L
LO
CH21 CH22
CH11 CH12 CH13 CH14 CH15 CH16
H L H L H L H L
CH17 CH18 CH19 CH20
SENSE
(OHMS, 4 WIRE)
INPUT
(V, 2 WIRE)
H L H L H L H L
CH7
CH8 CH9 CH10
H L H L H L H L
H L H L H L H L
INPUT SENSE CH1
CH2
CH3
CH4
CH5
CH6
CH1
CH11
Output HI
Output
LO
DC Voltage Calibrator
Note: Use shielded cables
to minimize noise.
Model 7700
Summary of Contents for 2701
Page 15: ...1 PerformanceVerification...
Page 47: ...2 Calibration...
Page 70: ...3 RoutineMaintenance...
Page 78: ...4 Troubleshooting...
Page 97: ...5 Disassembly...
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ...6 ReplaceableParts...
Page 125: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ...A Specifications...
Page 140: ...B CalibrationReference...