Model 2701 Service Manual
Troubleshooting
4-3
Power-on self-tests
RAM and EPROM tests
During the power-on sequence, the Model 2701 will perform a checksum test on its EPROM
and test its RAM. If the RAM tests fails, the instrument will lock up.
Flash memory tests
After the RAM and EPROM tests, the instrument will perform a checksum test on flash
memory. If a failure occurs, the unit will perform the comprehensive tests summarized in
Table 4-1
. If a failure occurs, an error message will be displayed, and the unit will automati-
cally enter the firmware upgrade mode (as indicated by “FW UPGRADE” message). Depend-
ing on the failure mode, firmware upgrades can be performed through the port(s) listed in the
table. You can also force the unit into the firmware upgrade mode by holding in the STEP key
during power-up.
Table 4-1
Flash memory failure modes
Flash memory
code sections
1
Status and error messages
Firmware upgrade port
3
FPGA
LOADING FPGA
FPGA CHKSUM
2
FPGA TIMEOUT
2
FPGA COMMERR
2
RS-232
Ethernet
LOAD ETHRNET
ETHER CHKSUM
2
ETHR TIMEOUT
2
RS-232
Main code
LOADING MAIN
MAIN CHKSUM
2
RS-232 or Ethernet
1. Code sections tested in sequence shown.
2. Error message displayed only on failure of specific test.
3. Firmware upgrade can be performed only through indicated port(s) depending on failure mode.
Summary of Contents for 2701
Page 15: ...1 PerformanceVerification...
Page 47: ...2 Calibration...
Page 70: ...3 RoutineMaintenance...
Page 78: ...4 Troubleshooting...
Page 97: ...5 Disassembly...
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ...6 ReplaceableParts...
Page 125: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ...A Specifications...
Page 140: ...B CalibrationReference...