Model 2701 Service Manual
Calibration Reference
B-5
:INIT
(:CALibration:PROTected:INITiate)
Purpose
To initiate comprehensive and factory calibration procedures.
Format
:cal:prot:init
Parameter
None
Description
The
:INIT
command enables Model 2701 calibration when performing
these procedures over the bus. This command must be sent to the unit after
sending the
:CODE
command, but before sending any other calibration
command.
NOTE
The
:INIT
command should be sent only once before performing either DC, AC, or
factory calibration. Do not send
:INIT
before each calibration step.
Example
:CAL:PROT:INIT
Initiate calibration.
:LOCK
(:CALibration:PROTected:LOCK)
Purpose
To lock out comprehensive or manufacturing calibration.
Format
:cal:prot:lock
Parameter
None
Description
The
:LOCK
command allows you to lock out both comprehensive and
manufacturing calibration after completing those procedures. Thus,
:LOCK
performs the opposite of enabling calibration with the
:CODE
command.
NOTE
To unlock comprehensive calibration, send the
:CODE
command. To unlock
manufacturing calibration, hold in the OPEN key while turning on the power.
Example
:CAL:PROT:LOCK
Lock out calibration.
Summary of Contents for 2701
Page 15: ...1 PerformanceVerification...
Page 47: ...2 Calibration...
Page 70: ...3 RoutineMaintenance...
Page 78: ...4 Troubleshooting...
Page 97: ...5 Disassembly...
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ...6 ReplaceableParts...
Page 125: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ...A Specifications...
Page 140: ...B CalibrationReference...