Model 2701 Service Manual
Calibration Reference
B-9
DC calibration commands
The :DC commands perform calibration of the DCV, DCI, and ohms functions.
Table B-2
summarizes these calibration commands along with parameter limits.
:STEP1
(:CALibration:PROTected:DC:STEP1)
Purpose
To perform front terminal short-circuit calibration.
Format
:cal:prot:dc:step1
Parameter
None
Description
:STEP1
performs the short-circuit calibration step in the comprehensive
calibration procedure. Connect a low-thermal short (Model 8610) to the
front panel input jacks before sending this command.
Example
:CAL:PROT:DC:STEP1
Perform short-circuit calibration.
Table B-2
DC calibration commands
Command
Description
Parameter limits
:CALibration
:PROTected
:DC
:STEP1
Front terminal short circuit.
:STEP2
Open circuit.
:STEP3 <NRf>
10V DC calibration step.
9 to 11
:STEP4 <NRf>
-10V DC calibration step.
-9 to -11
:STEP5 <NRf>
100V DC calibration step.
90 to 110
:STEP6 <NRf>
1k
Ω
4-wire calibration step.
900 to 1.1E3
:STEP7 <NRf>
10k
Ω
4-wire calibration step.
9E3 to 11E3
:STEP8 <NRf>
100k
Ω
4-wire calibration step.
90E3 to 110E3
:STEP9 <NRf>
1M
Ω
4-wire calibration step.
900E3 to 1.1E6
:STEP10 <NRf>
10mA DC calibration step.
9E-3 to 11E-3
:STEP11 <NRf>
100mA DC calibration step.
90E-3 to 110E-3
:STEP12 <NRf>
1A DC calibration step.
0.9 to 1.1
Summary of Contents for 2701
Page 15: ...1 PerformanceVerification...
Page 47: ...2 Calibration...
Page 70: ...3 RoutineMaintenance...
Page 78: ...4 Troubleshooting...
Page 97: ...5 Disassembly...
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ...6 ReplaceableParts...
Page 125: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ...A Specifications...
Page 140: ...B CalibrationReference...