1-12
Performance Verification
Model 2701 Service Manual
Verifying DC current
Check DC current accuracy by applying accurate DC currents from the DC current calibra-
tor to the AMPS input of the Model 2701 and verifying that the displayed readings fall within
specified limits.
Follow these steps to verify DC current accuracy:
1.
Connect the Model 2701 AMPS and INPUT LO jacks to the calibrator as shown in
Figure 1-3
. Be sure the INPUTS switch is in the FRONT position.
Figure 1-3
Connections for Model 2701 DC current verification
2.
Select the DC current measurement function by pressing the
DCI
key.
3.
Set the Model 2701 for the 20mA range.
4.
Source positive and negative full-scale currents for each of the ranges listed in
Table 1-4
and verify that the readings for each range are within stated limits.
Table 1-4
DCI limits
DCI range
Applied DC current*
Reading limits (1 year, 18°C to 28°C)
20mA
100mA
1A
3A
20.0000mA
100.0000mA
1.000000A
3.000000A**
19.98840 to 20.011160mA
99.8700 to 100.1300mA
0.999120 to 1.000880A
2.99628 to 3.00372A
* Source positive and negative currents with values shown.
** If the Fluke 5725 amplifier is not available, apply 2.2A from calibrator. Reading limits for 2.2A
input are: 2.197240 to 2.202760A.
!
Model 2701 Ethernet Multimeter / Data Acquisition System
RANGE
F
500V
PEAK
FRONT/REAR
3A 250V
AMPS
HI
INPUT
LO
SENSE
Ω
4 WIRE
INPUTS
350V
PEAK
1000V
PEAK
AUTO
SHIFT
LOCAL
POWER
RANGE
R
EXIT
ENTER
DIGITS RATE
REL
FILTER
TRIG
EX TRIG
STORE RECALL
OPEN
DCV
DCI
MATH
O U T P U T
RATIO
ACV
ACI
Ω
2
Ω
4
FREQ
TEMP
CH AVG
CONT
PERIOD SENSOR
LIMITS
ON/OFF
DELAY
HOLD
SAVE
SETUP
CONFIG
HALT
TYPE
LSYNC
TEST
MONITOR
STEP
SCAN
OCOMP
CH-OFF
CARD
CLOSE
Integra Series
RS-232
ETHERNET
Note: Be sure calibrator is set for
AMPS
Calibrator (Output DC Current)
normal current output.
Model 2701
INPUT LO
OUTPUT
HI
OUTPUT
LO
CAT I
Summary of Contents for 2701
Page 15: ...1 PerformanceVerification...
Page 47: ...2 Calibration...
Page 70: ...3 RoutineMaintenance...
Page 78: ...4 Troubleshooting...
Page 97: ...5 Disassembly...
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ...6 ReplaceableParts...
Page 125: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ...A Specifications...
Page 140: ...B CalibrationReference...