B-20
Calibration Reference
Model 2701 Service Manual
:LOCK
(:CALibration:PROTected:CARD1:LOCK)
Purpose
To lock out Model 7700 calibration.
Format
:cal:prot:card1:lock
Parameter
None
Description
The
:LOCK
command allows you to lock out Model 7700 calibration after
completing the procedure. Thus,
:LOCK
performs the opposite of enabling
calibration with the
:CODE
command.
Example
:CAL:PROT:CARD1:LOCK
Lock out card 1 calibration.
:LOCK?
(:CALibration:PROTected:CARD1:LOCK?)
Purpose
To read Model 7700 calibration lock status.
Format
:cal:prot:card1:lock?
Response
0
Calibration locked.
1
Calibration unlocked.
Description
The
:LOCK?
query requests status from the Model 2701 on Model 7700
calibration locked/unlocked state. Calibration must be enabled sending the
:CODE
command before calibration can be performed.
Example
:CAL:PROT:CARD1:LOCK?
Request card 1 cal lock state.
:RCOunt
(:CALibration:PROTected:CARD1:RCOunt)
Purpose
To reset card calibration count to 0.
Format
:cal:prot:card1:rco
Parameter
None
Description
The
:RCOunt
command resets the card calibration count reported by
:COUNt?
to 0.
Example
:CAL:PROT:CARD1:RCO
Rest card calibration count.
Summary of Contents for 2701
Page 15: ...1 PerformanceVerification...
Page 47: ...2 Calibration...
Page 70: ...3 RoutineMaintenance...
Page 78: ...4 Troubleshooting...
Page 97: ...5 Disassembly...
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ...6 ReplaceableParts...
Page 125: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ...A Specifications...
Page 140: ...B CalibrationReference...