Model 2701 Service Manual
Routine Maintenance
3-7
Plug-in module relay closure count
The Model 2701 keeps an internal count of the number of times each module relay has been
closed. This count will help you determine if and when any relays require replacement (see
module contact life specifications). The count can be read or reset only via remote as outlined
below.
Closure count commands
Table 3-2
summarizes closure count commands.
Reading relay closure count
To determine the closure count of specific channels, send this query via remote:
:ROUTe:CLOSe:COUNt? (@clist)
Here, clist is the summary of channels. For example, to determine the closure count of
channels 1 and 4 of a module in slot 1, the following query would be sent:
:ROUT:CLOS:COUN? (@101,104)
The following query would determine the closure count of slot 1 module channels
1 through 10:
:ROUT:CLOS:COUN? (@101:110)
Table 3-2
Closure count commands
Command
Description
:ROUTe
:CLOSe
:COUNt? (@clist)
:INTerval < NRf>
:INTerval?
:RCOunt (@clist)
Route subsystem.
Path to CLOSe commands.
Query count for channels in clist (channel list).
Set count update interval in minutes (1 to 1440).
Query count update interval.
Reset count for channels in clist.*
* Unit must be in manufacturing calibration mode. See text below.
Summary of Contents for 2701
Page 15: ...1 PerformanceVerification...
Page 47: ...2 Calibration...
Page 70: ...3 RoutineMaintenance...
Page 78: ...4 Troubleshooting...
Page 97: ...5 Disassembly...
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ...6 ReplaceableParts...
Page 125: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ...A Specifications...
Page 140: ...B CalibrationReference...