Model 2701 Service Manual
Performance Verification
1-15
6.
Source the nominal full-scale resistance values for the 100
Ω
-10M
Ω
ranges summarized
in
Table 1-6
and verify that the readings are within calculated limits.
7.
Connect the Model 2701 INPUT and SENSE jacks to the calibrator as shown in
Figure 1-6
.
8.
Disable external sense on the calibrator.
9.
Set the Model 2701 for the 100M
Ω
range.
10.
Source a nominal 100M
Ω
resistance value and verify that the reading is within calcu-
lated limits for the 100M
Ω
range.
Figure 1-6
Connections for Model 2701 resistance verification (100M
Ω
range)
Table 1-6
Limits for resistance verification
Ω
Range
Nominal
resistance
Nominal reading limits
(1 year, 18°C to 28°C)
Recalculated limits**
100
Ω∗
1k
Ω
10k
Ω
100k
Ω
1M
Ω
10M
Ω
100M
Ω
100
Ω
1k
Ω
10k
Ω
100k
Ω
1M
Ω
10M
Ω
100M
Ω
99.9880 to 100.0120
Ω
0.999894 to 1.000106k
Ω
9.99894 to 10.00106k
Ω
99.9890 to 100.0110k
Ω
0.999890 to 1.000110M
Ω
9.99590 to 10.00410M
Ω
99.7970 to 100.2030M
Ω
__________ to __________
Ω
__________ to __________ k
Ω
__________ to __________ k
Ω
__________ to __________ k
Ω
__________ to __________ M
Ω
__________ to __________ M
Ω
__________ to __________ M
Ω
* Enable O COMP (offset-compensated ohms) when testing 100
Ω
range.
** Calculate limits based on actual calibration resistance values and Model 2701 one-year resistance accuracy specifications. See
Verification limits.
!
Model 2701 Ethernet Multimeter / Data Acquisition System
RANGE
F
500V
PEAK
FRONT/REAR
3A 250V
AMPS
HI
INPUT
LO
SENSE
Ω
4 WIRE
INPUTS
350V
PEAK
1000V
PEAK
AUTO
SHIFT
LOCAL
POWER
RANGE
R
EXIT
ENTER
DIGITS RATE
REL
FILTER
TRIG
EX TRIG
STORE RECALL
OPEN
DCV
DCI
MATH
O U T P U T
RATIO
ACV
ACI
Ω
2
Ω
4
FREQ
TEMP
CH AVG
CONT
PERIOD SENSOR
LIMITS
ON/OFF
DELAY
HOLD
SAVE
SETUP
CONFIG
HALT
TYPE
LSYNC
TEST
MONITOR
STEP
SCAN
OCOMP
CH-OFF
CARD
CLOSE
Integra Series
RS-232
ETHERNET
Resistance Calibrator
Note: Use shielded, low-thermal cables to
minimize noise. Disable calibrator
external sense.
Model 2701
OUTPUT
HI
OUTPUT
LO
INPUT
HI
SENSE
HI
SENSE
LO
INPUT
HI
CAT I
Summary of Contents for 2701
Page 15: ...1 PerformanceVerification...
Page 47: ...2 Calibration...
Page 70: ...3 RoutineMaintenance...
Page 78: ...4 Troubleshooting...
Page 97: ...5 Disassembly...
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ...6 ReplaceableParts...
Page 125: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ...A Specifications...
Page 140: ...B CalibrationReference...