Model 2701 Service Manual
Calibration
2-17
AC voltage calibration
Follow these steps for AC voltage calibration:
1.
Connect the calibrator to the Model 2701 INPUT HI and LO terminals as shown in
Figure 2-4
.
2.
Perform the calibration steps summarized in
Table 2-11
. For each step:
•
Set the calibrator to the indicated voltage and frequency, make sure the unit is in
operate. (You must use the stated voltage and frequency.)
•
Send the indicated programming command.
•
Wait until the Model 2701 completes each step before continuing.
Table 2-11
AC voltage calibration programming steps
Calibration step
Calibrator voltage, frequency
Calibration command
10mV AC at 1kHz
100mV AC at 1kHz
100mV AC at 50kHz
1VAC at 1kHz
1VAC at 50kHz
10VAC at 1kHz
10VAC at 50kHz
100VAC at 1kHz
100VAC at 50kHz
700VAC at 1kHz
10.00000mV, 1kHz
100.0000mV, 1kHZ
100.0000mV, 50kHz
1.000000V, 1kHz
1.000000V, 50kHz
10.00000V, 1kHz
10.00000V, 50kHz
100.0000V, 1kHz
100.0000V, 50kHz
700.000V, 1kHz
:CAL:PROT:AC:STEP1
:CAL:PROT:AC:STEP2
:CAL:PROT:AC:STEP3
:CAL:PROT:AC:STEP4
:CAL:PROT:AC:STEP5
:CAL:PROT:AC:STEP6
:CAL:PROT:AC:STEP7
:CAL:PROT:AC:STEP8
:CAL:PROT:AC:STEP9
:CAL:PROT:AC:STEP10
Summary of Contents for 2701
Page 15: ...1 PerformanceVerification...
Page 47: ...2 Calibration...
Page 70: ...3 RoutineMaintenance...
Page 78: ...4 Troubleshooting...
Page 97: ...5 Disassembly...
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ...6 ReplaceableParts...
Page 125: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ...A Specifications...
Page 140: ...B CalibrationReference...