Model 2701 Service Manual
Calibration Reference
B-11
:STEP5
(:CALibration:PROTected:DC:STEP5)
Purpose
To program the 100V DC comprehensive calibration step.
Format
:cal:prot:dc:step5 <Cal_voltage>
Parameter
<Cal_voltage> = 90 to 110 [V]
Description
:STEP5
programs the 100V DC comprehensive calibration step. The
allowable range of the calibration voltage parameter is from 90 to 110, but
100 is recommended for best results.
Example
:CAL:PROT:DC:STEP5 100
Program 100V step.
:STEP6
(:CALibration:PROTected:DC:STEP6)
Purpose
To program the 1k
Ω
4-wire comprehensive calibration step.
Format
:cal:prot:dc:step6 <Cal_resistance>
Parameter
<Cal_resistance> = 900 to 1.1E3 [
Ω
]
Description
:STEP6
programs the 1k
Ω
4-wire resistance comprehensive calibration
step. The allowable range of the calibration resistance parameter is from 900
to 1.1E3, but 1E3 is recommended for best results.
Example
:CAL:PROT:DC:STEP6 1E3
Program 1k
Ω
step.
:STEP7
(:CALibration:PROTected:DC:STEP7)
Purpose
To program the 10k
Ω
4-wire comprehensive calibration step.
Format
:cal:prot:dc:step7 <Cal_resistance>
Parameter
<Cal_resistance> = 9E3 to 11E3 [
Ω
]
Description
:STEP7
programs the 10k
Ω
4-wire resistance comprehensive calibration
step. The allowable range of the calibration resistance parameter is from
9E3 to 11E3, but 10E3 is recommended for best results.
Example
:CAL:PROT:DC:STEP7 10E3
Program 10k
Ω
step.
Summary of Contents for 2701
Page 15: ...1 PerformanceVerification...
Page 47: ...2 Calibration...
Page 70: ...3 RoutineMaintenance...
Page 78: ...4 Troubleshooting...
Page 97: ...5 Disassembly...
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ...6 ReplaceableParts...
Page 125: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ...A Specifications...
Page 140: ...B CalibrationReference...