2701 Ethernet Multimeter/Data Acquisition System
DC CHARACTERISTICS
1
CONDITIONS: MED (1 PLC)
2
or 10 PLC or MED (1 PLC) with Digital Filter of 10
TEST
INPUT
ACCURACY: ±(ppm of r ppm of range)
CURRENT ±5%
RESISTANCE
(ppm = parts per million) (e.g., 10ppm = 0.001%)
TEMPERATURE
OR BURDEN
OR OPEN CKT.
24 Hour
4
90 Day
1 Year
COEFFICIENT
FUNCTION
RANGE
RESOLUTION
VOLTAGE
VOLTAGE
3
23°C±1°
23°C±5°
23°C±5°
0°–18°C & 28°–50°C
Voltage
11
100.0000 mV
0.1 µV
>10 G
Ω
15 + 30
25 + 35
30 + 35
(1 + 5)/°C
1.000000
V
1.0 µV
>10 G
Ω
15 + 6
25 + 7
30 + 7
(1 + 1)/°C
10.00000
V
10 µV
>10 G
Ω
10 + 4
20 + 5
30 + 5
(1 + 1)/°C
100.0000
V
100 µV
10 M
Ω
± 1%
15 + 6
35 + 9
45 + 9
(5 + 1)/°C
1000.000
V
5
1 mV
10 M
Ω
± 1%
20 + 6
35 + 9
50 + 9
(5 + 1)/°C
Resistance
6, 8
100.0000
Ω
100 µ
Ω
1 mA
6.9 V
20 + 20
80 + 20
100 + 20
(8 + 1)/°C
1.000000 k
Ω
1 m
Ω
1 mA
6.9 V
20 + 6
80 + 6
100 + 6
(8 + 1)/°C
10.00000 k
Ω
10 m
Ω
100 µA
6.9 V
20 + 6
80 + 6
100 + 6
(8 + 1)/°C
100.0000 k
Ω
100 m
Ω
10 µA
12.8 V
20 + 6
80 + 10
100 + 10
(8 + 1)/°C
1.000000 M
Ω
1.0
Ω
10 µA
12.8 V
20 + 6
80 + 10
100 + 10
(8 + 1)/°C
10.00000 M
Ω
7
10
Ω
0.7 µA // 10M
Ω
7.0 V
150 + 6
200 + 10
400 + 10
(70 + 1)/°C
100.0000 M
Ω
7
100
Ω
0.7 µA // 10M
Ω
7.0 V
800 + 30
2000 + 30
2000 + 30
(385 + 1)/°C
Continuity (2W)
21
1.000 k
Ω
100 m
Ω
1 mA
6.9 V
40 + 100
100 + 100
100 + 100
(8 + 1)/°C
Current
20.00000 mA
10 nA
< 0.2
V
60 + 30
300 + 80
500 + 80
(50 + 5)/°C
100.0000 mA
100 nA
< 0.1
V
100 + 300
300 + 800
500 + 800
(50 + 50)/°C
1.000000
A
1.0 µA
< 0.5
V
9
200 + 30
500 + 80
800 + 80
(50 + 5)/°C
3.000000
A
10 µA
< 1.5
V
9
1000 + 15
1200 + 40
1200 + 40
(50 + 5)/°C
Channel (Ratio)
10
Ratio Accuracy = Accuracy of selected Channel Range + Accuracy of Paired Channel Range
Channel (Average)
10
Average Accuracy = Accuracy of selected Channel Range + Accuracy of Paired Channel Range
TEMPERATURE
19
(Displayed in °C, °F, or K. Exclusive of probe errors.)
Thermocouples (Accuracy based on ITS-90.)
90 Day/1 Year (23°C ± 5°C)
Relative to
Using
Temperature
Simulated
77xx
Coefficient
Type
Range
Resolution Reference Junction
Module
0°–18°C & 28°–50°C
J
–200 to +760 °C
0.001°C
0.2°C
1.0°C
0.03°C/°C
K
–200 to +1372°C
0.001°C
0.2°C
1.0°C
0.03°C/°C
N
–200 to +1300°C
0.001°C
0.2°C
1.0°C
0.03°C/°C
T
–200 to +400°C
0.001°C
0.2°C
1.0°C
0.03°C/°C
E
–200 to +1000°C
0.001°C
0.2°C
1.0°C
0.03°C/°C
R
0 to +1768°C
0.1
°C
0.6°C
1.8°C
0.03°C/°C
S
0 to +1768°C
0.1
°C
0.6°C
1.8°C
0.03°C/°C
B
+350 to +1820°C
0.1
°C
0.6°C
1.8°C
0.03°C/°C
4-Wire RTD:
(100
Ω
platinum [PT100], D100, F100, PT385, PT3916, or user type. Offset compensation On)
–200° to
630°C
0.01 °C
0.06°C
0.003°C/°C
Thermistor: (2.2k
Ω
, 5k
Ω
, and 10k
Ω
)
20
–80° to
150°C
0.01 °C
0.08°C
0.002°C/°C
DC SYSTEM SPEEDS
15,18
RANGE CHANGES
16
:
50/s (42/s) (excludes 4 Wire Ohms).
FUNCTION CHANGES
16
:
50/s (42/s).
AUTORANGE TIME
16
:
<30ms.
ASCII READINGS TO RS-232 (115.2k BAUD):
300/s.
MAX. EXTERNAL TRIGGER RATE:
2000/s.
DC MEASUREMENT SPEEDS
15
Single Channel, 60Hz (50Hz) Operation
FUNCTION
DIGITS
READINGS/s
PLCs
DCV, DCI,
Ω
(<10M),
6.5
12,16
5
(4)
10
Thermocouple,
6.5
16
35
(28)
1
Thermistor
6.5
12,16
50
(48)
1
5.5
12,16
150
(120)
0.1
5.5
16, 17
300
(240)
0.1
5.5
17
500
(400)
0.1
4.5
17
3000 (3000)
0.006
3.5
17
3500 (3500)
0.002
4W
Ω
(<10M)
6.5
16
1.4
(1.1)
10
6.5
16
15
(12)
1
5.5
16, 17
33
(25)
0.1
4W
Ω
OComp, RTD
6.5
16, 17
0.9
(0.7)
10
6.5
16, 17
8
(6.4)
1
5.5
16, 17
18 (14.4)
0.1
Channel (Ratio),
6.5
16, 17
2.5
(2)
10
Channel (AVG)
6.5
16, 17
15
(12)
1
5.5
16, 17
25
(20)
0.1
Multiple Channels Into Memory
15, 18
7710 Scanning DCV
500/s
7703 Scanning DCV
390/s
7703 Scanning DCV with Limits or Time Stamp On
390/s
7703 Scanning DCV alternating 2W
Ω
95/s
Multiple Channels, Into and Out of Memory
15, 16, 17, 18
Channels/s
7710 Scanning DCV
440/s
7703 Scanning DCV
360/s
7703 Scanning DCV with Limits or Time Stamp On
360/s
7703 Scanning DCV alternating 2W
Ω
90/s
7702 Scanning DCV
75/s
7700 and 7708 Scanning Temperature (T/C)
60/s
SPEC-2701
Summary of Contents for 2701
Page 15: ...1 PerformanceVerification...
Page 47: ...2 Calibration...
Page 70: ...3 RoutineMaintenance...
Page 78: ...4 Troubleshooting...
Page 97: ...5 Disassembly...
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ...6 ReplaceableParts...
Page 125: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ...A Specifications...
Page 140: ...B CalibrationReference...