Model 2701 Service Manual
Calibration Reference
B-3
Command
Description
:CALibration
:PROTected
:DC
:STEP9 <NRf>
1M
Ω
4-wire step.
:STEP10 <NRf>
10mA DC step.
:STEP11 <NRf>
100mA DC step.
:STEP12 <NRf>
1A DC step.
:AC
AC cal steps.
:STEP1
10mV AC at 1kHz step.
:STEP2
100mV AC at 1kHz step.
:STEP3
100mV AC at 50kHz step.
:STEP4
1V AC at 1kHz step.
:STEP5
1V AC at 50kHz step.
:STEP6
10V AC at 1kHz step.
:STEP7
10V AC at 50kHz step.
:STEP8
100V AC at 1kHz step.
:STEP9
100V AC at 50kHz step.
:STEP10
700V AC at 1kHz step.
:STEP11
100mA AC at 1kHz step.
:STEP12
1A AC at 1kHz step.
:STEP13
2A AC at 1kHz step.
:STEP14
1V AC at 3Hz step
1
.
:STEP15
1V AC at 1kHz step
1
.
NOTE: Upper-case letters indicated short form of each command. For example, instead of sending
“:CALibration:PROTected:INITiate,” you can send “:CAL:PROT:INIT.”
1
DC:STEP0, AC:STEP14, and AC:STEP15 are one-time factory calibration points and are valid only in manufacturing calibration
mode.
Table B-1 (Continued)
Remote calibration command summary
Summary of Contents for 2701
Page 15: ...1 PerformanceVerification...
Page 47: ...2 Calibration...
Page 70: ...3 RoutineMaintenance...
Page 78: ...4 Troubleshooting...
Page 97: ...5 Disassembly...
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ...6 ReplaceableParts...
Page 125: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ...A Specifications...
Page 140: ...B CalibrationReference...