Model 2701 Service Manual
Calibration Reference
B-17
Model 7700 calibration commands
Table B-4
summarizes calibration commands for the Model 7700 plug-in module. Note that
CARD1 commands calibrate the card in Slot 1, while CARD2 commands request calibration
count and date information from a card in Slot 2.
NOTE
A Model 7700 must be installed in Slot 1 through a Model 7797 Calibration System
card to be calibrated. See “Model 7700 calibration” in Section 2.
Table B-4
Model 7700 calibration commands
Command
Description
:CALibration
Calibration root command.
:PROTected
All commands in this subsystem are protected by the calibration lock
(except queries and :CODE).
:CODE '<up to 8 char. string>'
Send calibration code. (Default KI002701.)
:CARD1
Path to card in slot 1.
:INITiate
Initiate card calibration.
:COUNt?
Request number of times card had been calibrated.
:RCOunt
Reset card calibration count to 0.
:DATE?
Request card cal date.
:SAVE
Save cal constants to card EEPROM.
:LOCK
Lock out calibration.
:LOCK?
Request cal lock state. (0 = locked, 1 = unlocked.)
:STEP0 <NRf>
Temperature sensor cold calibration (<NRf> = temperature, ˚C).
:CARD2
Path to card in slot 2.
:COUNt?
Request number of times card has been calibrated.
:DATE?
Request card calibration date.
Summary of Contents for 2701
Page 15: ...1 PerformanceVerification...
Page 47: ...2 Calibration...
Page 70: ...3 RoutineMaintenance...
Page 78: ...4 Troubleshooting...
Page 97: ...5 Disassembly...
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ...6 ReplaceableParts...
Page 125: ......
Page 126: ......
Page 127: ......
Page 128: ......
Page 129: ......
Page 130: ......
Page 131: ...A Specifications...
Page 140: ...B CalibrationReference...