Operating Instructions

Model MM47

Digital Multimeter

SAFETY INFORMATION

The following safety information must be observed
to ensure maximum personal safety during the 
operation of this meter:
Do not use the meter if the meter or test leads look 
damaged, or if you suspect that the meter is not
operating properly.
Never ground yourself when taking electrical 
measurements. Do not touch exposed metal pipes, 
outlets, fixtures, etc, which might be at ground 
potential. Keep your body isolated from ground by
using dry clothing, rubber shoes, rubber mats, or
any insulating material. 
Turn off power to the circuit under test before 
cutting, unsoldering, or breaking the circuit. Small
amounts of current could be dangerous. 
Use caution when working above 60V DC or 30V AC
rms. Such voltages pose a shock hazard.
When using the probes, keep your fingers behind
the finger guards on the probes. 

MM47 REV3 7/05

NOTE

This instrument is only to be used by a suitably
trained and competent person.

The symbols used on this instrument are:

Measuring voltage which exceeds the limits of the
multimeter may damage the meter and expose the
operator to a shock hazard. Always recognise the
meter voltage limits as stated on the front of the
meter. 

SPECIFICATIONS
Display: 
3½   digit liquid crystal display (LCD) with a 
maximum reading of 1999.  
Polarity: Automatic, positive implied, negative
polarity indication. 
Overrange: (OL) or (-OL) is displayed.
Zero: automatic.

Low Battery Indication: The 

is displayed when

the battery voltage drops below the operating level.  
Measurement rate: 2.5 times per second, nominal.
Operating environment: 0°C to 50°C at < 70% R.H.
Storage temperature: -20°C to 60°C, 0 to 80% R.H.
with battery removed from meter.  
Accuracy: Stated accuracy at 23°C ± 5°C, < 75% R.H.
Battery life: 200 hours typical with carbon-zinc.
Battery: Single standard 9-volt battery, NEDA 1604,
JIS 006P, IEC 6F22.
Dimensions: 147mm (H)x 70mm (W)x 39mm (D).
Weight: Approx. 340g including holster.
Accessories: One set test leads, 9V battery
(installed), one thermocouple probe and Operating
Instructions. 

D C VOLTS
Ranges: 
200mV, 2V, 20V, 600V
Resolution: 100µV
Accuracy: ± (1.2% rdg + 1dgt)
Input impedance: 10M

Ω

Overload protection: 600VDC or AC rms

AC VOLTS: (50Hz - 500Hz) 
Ranges: 200mV, 2V, 600V
Resolution: 1mV
Accuracy: 200mV, 2V ranges ± (1.5% rdg + 4dgts)
600V range ± (2.0% rdg + 4dgts)
Input impedance: 4.5M

Ω

Overload protection: 600V DC or AC rms

RESISTANCE
Ranges: 
200

Ω

, 2k

Ω

, 20k

Ω

, 200k

Ω

, 2000k

Ω

, 20M

Ω

,

200M

Ω, 

2000M

Ω.

Resolution: 100m

Ω

Accuracy: ± (1.0% rdg + 4 dgts) on 200

Ω

ranges to

2000k

Ω

± (2.0% rdg + 4 dgts) on 20M

Ω

range

± [(5.0% rdg - 10 dgts) + 10 dgts] on 2000M

Ω

range
Open Circuit volts: 0.3V dc typical (3.0V DC on
200

Ω

, 2000M

Ω

ranges).

Overload Protection: 500V DC or AC rms

CONTINUITY
Audible indication: 
Less than 100

Ω

Response Time: 100mS

Overload protection: 500V DC or AC rms

DIODE TEST 
Test current: 
1.0mA ± 0.6mA
Accuracy: ± (3.0% rdg + 1 dgt)
Open circuit volts: 3.0Vdc typical
Overload protection: 500V DC or AC rms

CAPACITANCE
Ranges: 
20µF, 200µF, 2000µF, 20mF
Accuracy: ± (4.0% rdg + 10 dgts) on all ranges
Test frequency: 21Hz
Test voltage: <3.5V
Input protection: 0.1A/250V fast acting fuse

Caution, risk of electric shock

Caution, refer to accompanying
documents

Equipment protected throughout
by Double Insulation (Class II)

Alternating current

Direct current

Ground

z

~

HIGH ENERGY SYSTEMS

GS38 advises fused test leads to be used when 

working on high energy systems. Suitable leads are

Martindale TL55 (or TL58 for 3 phase use).

The instrument meets the requirements for double
insulation to BS EN 61010-1 at 600V Installation
Category

III

Installation Category

III

(distribution

level) relates to transient overvoltage likely to be
found in fixed installation wiring.